GREISINGER ResOx 5695 Operating Manual Download Page 20

H84.0.15.6C-01 

Operating manual ResOx 5695 

Page 20 of 22 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

18 Error and System Messages 

Display 

Meaning 

Remedy 

 

low battery voltage, device will continue to work 
for a short time 

replace battery 

If mains operation: wrong voltage 

replace power supply, if fault continues to 
exist: device damaged 

 

low battery voltage 

replace battery 

If mains operation: wrong voltage 

Check/replace power supply, if fault 
continues to exist: device damaged 

No display 

or 

weird display 

 

Device does not 

react on keys 

low battery voltage 

replace battery 

If mains operation: wrong voltage 

check/replace power supply, if fault 
continues to exist: device damaged 

system error 

disconnect battery or power supply, wait 
some time, re-connect 

device defective 

return to manufacturer for repair 

SENS 

ERRO

 

sensor error: no sensor cable connected 

connect suitable sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.1 

Value exceeding measuring range 

Check: Is the value exceeding the specified 
measuring range? ->value too high! 

Wrong sensor connected 

Check sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.2 

Value below display range 

Check: Is the value below the specified 
measuring range? ->value too low! 

Wrong sensor connected 

Check sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.7 

system error 

return to manufacturer for repair 

If 

BAT

 is flashing, the battery will be exhausted soon. Further measurements are possible for short time. 

If 

BAT

 is displayed continuously the battery is ultimately exhausted and has to be replaced. Further 

measure

ments aren’t possible any more.

 

Messages During Calibration/Adjustment 

>CAL< 

CAL flashing in 

display 

either preset calibration interval has expired or 
last calibration is not valid 

device has to be calibrated! 

[AL  ERR.1

 

wrong reference point at air  

check sensor and reference gas 

[AL  ERR.2

 

slope too low 

 

reference gas wrong 

check sensor and reference gas 

sensor element is defect  

replace sensor element  

[AL  ERR.3

 

slope too high 

 

reference gas wrong 

check sensor and reference gas 

s

ensor element is defect  

replace sensor element  

[AL  ERR.4

 

incorrect calibration temperature 

calibration can only be done at 0…50 °C 

[AL  ERR.5

 

Zero value to low/negative 

 

sensor element is defect  

replace sensor element  

[AL  ERR.6

 

zero value to high 

 

reference gas wrong 

check sensor and reference gas 

s

ensor element is defect  

replace sensor element  

[AL  ERR.7

 

incorrect calibration pressure 

check calibration pressure 

[AL  ERR.8

 

signal not stable / timeout 

check sensor and reference gas 

[AL  ERR.9

 

sensor not known: cannot be calibrated 

check sensor and wiring 

Summary of Contents for ResOx 5695

Page 1: ...ger de www greisinger de Operating manual Water Proof oxygen meter for oxygen in gases with Integrated temperature and pressure measuring and data logger As of version 1 0 ResOx 5695 Please carefully...

Page 2: ...ENTS 11 7 OXYGEN MEASURING IN GASES PLEASE NOTE 11 7 1 APPLICATION OF THE DIFFERENT SENSOR TYPES 11 8 EXCHANGING THE SENSOR ELEMENT 12 9 CALIBRATION OF THE OXYGEN SENSOR 13 9 1 ONE POINT CALIBRATION A...

Page 3: ...centration in Vol O2 For the measuring an external sensor of the type GOO GGO or GGA is necessary The measuring take place at the extern opening Sensor Due to the properties of the sensor it has to be...

Page 4: ...ed to when using the device Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under 20 Speci...

Page 5: ...s operation When connecting a mains cable or USB interface cable please take care to connect only allowed components The output voltage of a connected power supply unit has to be between 4 5 and 5 5 V...

Page 6: ...ration Confirm settings return to measuring Store Enter Measuring with Auto Hold off Hold and save current measuring value HLD is displayed with Auto Hold on Start new measuring It is finished when HL...

Page 7: ...n to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or atta...

Page 8: ...l configuration 9 12 H 2 P 02 hPa Oxygen partial pressure display in hPa P 02 mmHg Oxygen partial pressure display in mmHg RES K Best O2 resolution LO Low O2 resolution calm value display standard LCD...

Page 9: ...pressure measuring is shifted for the entered value This can be used to compensate sensor deviations OFF No zero adjustment for pressure measurement 0 hPa SET AL Set Alarm Configuration of the alarm...

Page 10: ...The sensor state evaluation will be stored after a successful calibration of the oxygen sensor 6 1 2 Mounting operation position The optimum position for the sensor membrane is to point downwards If...

Page 11: ...have to be flushed with clean air or nitrogen Integrated in GGO 581 GGA 581 GOO 581 Recommended application areas 0 100 Vol O2 Preferred of measurements with extremely small oxygen content e g 0 3 Vol...

Page 12: ...ynamic pressure due to gas flow this type is optimal It can be screwed tightly into systems with small over or under pressure Attention Mind the maximum pressure The instrument pressure port is connec...

Page 13: ...e which should be applied 0 2 for pure oxygen NULL for 0 oxygen e g pure nitrogen As long as the display blinks the instrument recognises no valid reference As soon the values for oxygen and temperatu...

Page 14: ...ate of the recording For the evaluation of the data the software GSOFT3050 version V3 0 or higher has to be used The software also allows easy configuration and starting of the logger When the logger...

Page 15: ...lection After that press again automatic recording is starting Whenever a measuring is recorded the display shows St XXXXX XXXXX is the number of the measuring If the logger memory is full the display...

Page 16: ...ure correct sensor calibration adjustment The device provides the following functions to help with this 12 1 Calibration Adjustment interval INT You can input the interval after which the device remin...

Page 17: ...es battery life time If the device is used together with interface adapter USB 5100 the device is supplied from the interface device pin assignment 1 external supply 5V 50mA 2 GND 3 TxD RxD 3 3V Logic...

Page 18: ...f display x x x x 176 read min measuring range x 208 read channel count x x x x 177 read max measuring range x 222 read turn off delay x x x x 178 read measuring range unit x 223 Set turn off delay x...

Page 19: ...ress of water and dust may be lost Avoid unnecessary opening of the instrument 1 Open the 3 Phillips screws at the backside of the instrument 2 Lay down the still closed instrument so that the display...

Page 20: ...ay range Check Is the value below the specified measuring range value too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error...

Page 21: ...ressure Lo 0 1100 hPa O2 Hi 0 0 1100 0 hPa O2 Sensor temperature 5 0 50 0 C NTC 10k integrated in GGO GOO GGA cable Absolute pressure 300 5000 hPa abs integrated pressure sensor with pressure port Acc...

Page 22: ...nction Device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off delay The power off delay can be set to values between 1and 1...

Reviews: