GREISINGER ResOx 5695 Operating Manual Download Page 11

H84.0.15.6C-01 

Operating manual ResOx 5695 

Page 11 of 22 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

6.2  Sensor elements 

 

Unsuitable for underwater-diving-application e.g. Rebreather. 

6.2.1  GOEL 370 acidic electrolyte 

 

Integrated in GGO 570, GGA 570, GOO 570.  
Recommended application areas 0...35 Vol. % O

2

 (above, the measuring accuracy is 

reduced). 
Also suitable for gases with high CO

2

-proportion or for CO

2

-gas. The effect of the 

acidic electrolyte is that the sensor is resistant against the CO

2

-gas and still hold the 

stability. 

6.2.2  GOEL 381 basic electrolyte 

 

Short-time exposition of up to 10% CO

2

 is not problematic (for example 15 minutes. 

up to 10 times per day) for the GOEL 380 (e.g. exhaust measuring). If there is 
measured more often with elevated CO

2

-concentration or at CO

2

-concentrations 

above 10%, the exposition time has to be kept as short as possible and sufficient 
measuring breaks should be made. 
If the sensor is not exposed to free air during measuring pauses, the connected 
tubes etc. have to be flushed with clean air or nitrogen. 

 

Integrated in GGO 581, GGA 581, GOO 581. 
Recommended application areas 0...100 Vol. % O

2

.  

Preferred of measurements with extremely small oxygen content (e.g. < 0.3 Vol. % 
O

2

), e.g. protective atmosphere, or more than 35 % Vol O

2.

 

For gases without larger CO2 concentration 

Oxygen measuring in gases- please note 

The instrument is designed for measuring the oxygen partial pressure or the oxygen concentration (% Vol. 
O

2

) calculated from partial pressure and ambient pressure) in gases. Please keep in mind. 

The sensor consists of a sensing element (GOEL 3xx) enclosed in a sensor housing (GGO/ GGA/ GOO). 
When purchasing a Sensor GGO/GGA/GOO 5xx, a sensor element is already integrated,  
e.g. a GGO 570: contains housing GGO and a sensor element GOEL 370 

7.1  Application of the different sensor types 

7.1.1  GGO housing (closed sensor) 

 For measurements at atmosphere

 and in systems without over or under pressure, the GGO... is sufficient. 

Additionally the GGO can be screwed tightly into systems with small 
over or under pressure.  
Attention! Mind the maximum pressure and the maximum pressure 
difference at the membrane. 
If instrument and sensor pressure are different, please connect the 
pressure port of instrument to measuring pressure, otherwise it will 

be compensated wrong! 
 

 

Summary of Contents for ResOx 5695

Page 1: ...ger de www greisinger de Operating manual Water Proof oxygen meter for oxygen in gases with Integrated temperature and pressure measuring and data logger As of version 1 0 ResOx 5695 Please carefully...

Page 2: ...ENTS 11 7 OXYGEN MEASURING IN GASES PLEASE NOTE 11 7 1 APPLICATION OF THE DIFFERENT SENSOR TYPES 11 8 EXCHANGING THE SENSOR ELEMENT 12 9 CALIBRATION OF THE OXYGEN SENSOR 13 9 1 ONE POINT CALIBRATION A...

Page 3: ...centration in Vol O2 For the measuring an external sensor of the type GOO GGO or GGA is necessary The measuring take place at the extern opening Sensor Due to the properties of the sensor it has to be...

Page 4: ...ed to when using the device Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under 20 Speci...

Page 5: ...s operation When connecting a mains cable or USB interface cable please take care to connect only allowed components The output voltage of a connected power supply unit has to be between 4 5 and 5 5 V...

Page 6: ...ration Confirm settings return to measuring Store Enter Measuring with Auto Hold off Hold and save current measuring value HLD is displayed with Auto Hold on Start new measuring It is finished when HL...

Page 7: ...n to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or atta...

Page 8: ...l configuration 9 12 H 2 P 02 hPa Oxygen partial pressure display in hPa P 02 mmHg Oxygen partial pressure display in mmHg RES K Best O2 resolution LO Low O2 resolution calm value display standard LCD...

Page 9: ...pressure measuring is shifted for the entered value This can be used to compensate sensor deviations OFF No zero adjustment for pressure measurement 0 hPa SET AL Set Alarm Configuration of the alarm...

Page 10: ...The sensor state evaluation will be stored after a successful calibration of the oxygen sensor 6 1 2 Mounting operation position The optimum position for the sensor membrane is to point downwards If...

Page 11: ...have to be flushed with clean air or nitrogen Integrated in GGO 581 GGA 581 GOO 581 Recommended application areas 0 100 Vol O2 Preferred of measurements with extremely small oxygen content e g 0 3 Vol...

Page 12: ...ynamic pressure due to gas flow this type is optimal It can be screwed tightly into systems with small over or under pressure Attention Mind the maximum pressure The instrument pressure port is connec...

Page 13: ...e which should be applied 0 2 for pure oxygen NULL for 0 oxygen e g pure nitrogen As long as the display blinks the instrument recognises no valid reference As soon the values for oxygen and temperatu...

Page 14: ...ate of the recording For the evaluation of the data the software GSOFT3050 version V3 0 or higher has to be used The software also allows easy configuration and starting of the logger When the logger...

Page 15: ...lection After that press again automatic recording is starting Whenever a measuring is recorded the display shows St XXXXX XXXXX is the number of the measuring If the logger memory is full the display...

Page 16: ...ure correct sensor calibration adjustment The device provides the following functions to help with this 12 1 Calibration Adjustment interval INT You can input the interval after which the device remin...

Page 17: ...es battery life time If the device is used together with interface adapter USB 5100 the device is supplied from the interface device pin assignment 1 external supply 5V 50mA 2 GND 3 TxD RxD 3 3V Logic...

Page 18: ...f display x x x x 176 read min measuring range x 208 read channel count x x x x 177 read max measuring range x 222 read turn off delay x x x x 178 read measuring range unit x 223 Set turn off delay x...

Page 19: ...ress of water and dust may be lost Avoid unnecessary opening of the instrument 1 Open the 3 Phillips screws at the backside of the instrument 2 Lay down the still closed instrument so that the display...

Page 20: ...ay range Check Is the value below the specified measuring range value too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error...

Page 21: ...ressure Lo 0 1100 hPa O2 Hi 0 0 1100 0 hPa O2 Sensor temperature 5 0 50 0 C NTC 10k integrated in GGO GOO GGA cable Absolute pressure 300 5000 hPa abs integrated pressure sensor with pressure port Acc...

Page 22: ...nction Device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off delay The power off delay can be set to values between 1and 1...

Reviews: