T20.0.4X.6C-03 

 Operating Manual GTMU - MP 

page 4 of 6 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

8  Display Functions 

8.1  Measuring display 

During normal operation the 

temperature

 in 

[°C]

 or 

[°F] 

is displayed. 

 

 

 

8.2  Min/Max Value Memory 

watch Min values (Lo):  press 

▼ shortly once 

display changes between „Lo„ and Min values 

watch Max values (Hi):  press 

▲ shortly once  

display changes between „Hi„ and Max values  

restore current values:  press 

▼ or ▲ once again 

current values are displayed 

clear Min-values:  

press 

▼ for 2 seconds 

Min values are cleared. Display shows 

shortly „CLr„. 

clear Max-values:  

press 

▲ for 2 seconds 

Max values are cleared. Display shows shortly „CLr„. 

After 10 seconds the currently measured values will be displayed again. 
 

9  Error and system messages 

Display 

Description 

Possible fault cause 

Remedy 

Err.1 

Measuring range 
exceeded 

Wrong signal 

Temperatures above the measuring 
range are not allowed. 

Err.2

 

Measuring value below 
measuring range 

Wrong signal 

Temperatures below the measuring 
range are not allowed. 

Err.7 

System fault 

Error in device 

Disconnect from supply and reconnect. 
If error remains: return to manufacturer 

Err.9 

Sensor error 

Sensor or cable defective 

Check sensors, cable and connections, 
damages visible? 

Er.11 

Calculation not possible 

Calculation variable 
missing or invalid 

Check temperature 

8.8.8.8 

Segment test 

The transducer performs a display test for 2 seconds after power 
up. After that it will change to the display of the measuring. 

 
 

 