background image

10 | Technical data

G 1720

22 / 25

B-H86.0.0C.DB2-1.0

10 Technical data

Measuring range temperature

-70.0 .. +250.0 °C (-94.0 .. +482.0 °F)

Accuracy temperature

-20 .. +100 °C: ±0.1 K ±1 digit

otherwise: ±0.2 % of measured value ± 2 digits

Response time T90 water (0.4 m/s)

approx. 3 s

Measuring cycle

approx. 2 measurements per second

Temperature connections

Permanently connected Pt1000 sensor (EN 60751)

Display

3-line segment LCD, additional symbols, illuminated (adjustable
white, permanent illumination)

Additional functions

Min/max/hold, alarm (optical and acoustic)

Compensation

Offset and gradient correction

Housing

Break-proof ABS housing

Protection rating

IP65 / IP67

Dimensions L*W*H
[mm]

108 * 54 * 28 mm without kink protection

Weight

150 g, incl. battery and sensor

Operating conditions

-20 to 50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.)

Storage temperature

-20 to 70 °C

Current supply

2*AA battery (included in the scope of delivery)

Current requirement/

battery life

approx. 0.4 mA, approx. 2 mA with lighting

Service life > 5000 hours with alkaline batteries (without backlight-
ing)

Battery indicator

4-stage battery status indicator,

Replacement indicator for depleted batteries: "BAT"

Auto-power-OFF function

The device switches off automatically if this is activated

Directives and standards

The devices conform to the following Directives of the Council for
the harmonisation of legal regulations of the Member States:

2014/30/EU EMC Directive

2011/65/EU RoHS

Applied harmonised standards:

EN 61326-1:2013 Emission limits: Class B
Immunity according to Table 2
Additional errors: < 0.5 % FS

EN 50581:2012

Summary of Contents for G 1720

Page 1: ...Operating manual G 1720 Alarm thermometer Permanently connected sensor Waterproof Precise and fast EN...

Page 2: ...ments 10 5 Operation 11 5 1 Commissioning 11 5 1 1 Explanation 11 5 2 Configuration 11 5 2 1 Explanation 11 5 2 2 Opening the configuration menu 11 5 2 3 Configuring parameters of the configuration me...

Page 3: ...Table of contents B H86 0 0C DB2 1 0 3 25 9 Disposal 21 10 Technical data 22 11 Spare parts and accessories 23 12 Service 24 12 1 Manufacturer 24 12 2 Repairs 24 12 3 Sales subsidiaries 25...

Page 4: ...Legal notices The liability and warranty of the manufacturer for damages and consequential dam ages are voided with misuse disregarding this operating manual disregarding safety notices assignment of...

Page 5: ...duct Outcome of an action Result consequence or effect of an instruction Emphases In order to simplify legibility and provide a clearer overview various sections informa tion are emphasised 1234 Displ...

Page 6: ...can only be guaranteed if generally applicable safety precautions and the device specific safety instructions for this document are observed If these notices are disregarded personal injury or death...

Page 7: ...le insertion probes with care Fit a protective cap on the measuring probe NOTE If the product is stored at a temperature above 50 C or is not used for an extended period of time the batteries must be...

Page 8: ...which purpose this document makes a valuable contribution The instructions in this docu ment must be understood observed and followed In order to ensure that no risks arise from the interpretation of...

Page 9: ...l impressive features include the dust proof and waterproof design in accord ance with IP 65 67 and the 3 line illuminated display which offers overhead display at the push of a button The product can...

Page 10: ...easurement of the current temperature in min max hold mode with unit 4 3 Operating elements On Off button Press briefly Switch on the product Activate deactivate lighting Long press Switch off the pro...

Page 11: ...roduct for your purposes NOTE There are various configuration parameters available depending on the product ver sion and configuration They can differ depending on the product version and configur ati...

Page 12: ...d parameter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up button and Down button 3 The available configuration options are listed for each parame...

Page 13: ...ted with the zero point correction and the gradient correction If an adjustment is made you change the pre adjusted factory settings This is signalled with the T OF or T SL when the product is switche...

Page 14: ...ired parameter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up button and Down button 3 The available configuration options are listed for each par...

Page 15: ...o point T OF 0 2 C 0 0 C 0 2 C T SL setpoint gradient correction display gradient correction T OF 1 100 T SL 37 0 C 36 6 C 0 2 1 100 0 54 Outcome of an action The changed value is saved and the Config...

Page 16: ...ntal temperature influ ence the measurement result NOTE Thermally conductive paste between the measuring sensor and surface can also in crease measurement accuracy in some cases 6 1 3 Cooling evaporat...

Page 17: ...ate for a certain length of time If the BAT display text appears in the main display the battery voltage is no longer ad equate for operation of the product Now the battery is fully depleted 7 2 2 Cha...

Page 18: ...are categorised as ISO calibration certificates and DAkkS calibration certificates The purpose of the calibration is to verify the precision of the measuring device by comparing it with a traceable re...

Page 19: ...0 0C DB2 1 0 19 25 NOTE The product is delivered with a test report This confirms that the measuring device has been adjusted and tested NOTE Only the manufacturer can check the basic settings and mak...

Page 20: ...m error Product is defective Battery depleted Error in the product Product is defective Replace battery Send in for repair BAT Battery depleted Battery depleted Replace battery ERR 1 Measuring range e...

Page 21: ...se online at www ghm group de and sent it in with the product NOTE The device must not be disposed of with household waste If the product is disposed of please take it to a municipal collection point...

Page 22: ...ut kink protection Weight 150 g incl battery and sensor Operating conditions 20 to 50 C 0 to 95 r h temporarily 100 r h Storage temperature 20 to 70 C Current supply 2 AA battery included in the scope...

Page 23: ...ed below Article Number Name Description 610049 Mignon battery AA Mignon AA spare battery 611373 ST G1000 Device protection bag with 1 round cut out A complete list of all accessories and spare parts...

Page 24: ...nger de WEEE reg no DE 93889386 12 2 Repairs Defective products are repaired professionally and quickly in our service centre Open hours and contact Monday to Thursday from 8 00 to 16 00 Friday from 8...

Page 25: ...UP France SAS Parc des Pivolles 9 Rue de Catalogne 69150 D cines Charpieu Lyon FRANCE Phone 33 4 72 37 45 30 a jouanilou ghm group fr India GHM Messtechnik India Pvt Ltd 209 Udyog Bhavan Sonowala Road...

Reviews: