7 | Error and system messages
G 1400 series
18 / 22
B-H87.0.0X.DK2-3.0
7
Error and system messages
Display
Meaning
Possible
causes
Remedy
----
Range switches or
measured value
unstable
Impurities or air
bubbles
Measured value far
outside of the
measuring range
Controlling instable
Contamination or air
bubbles on cell
Measurement outside
permissible range
Measuring cell defect
Wait for the transient
effect of the controller
clean contamination /
bubbles
Keep measurement within
the permissible range
Send in for repair
SENS ERRO
Sensor cable
defect
Sensor or probe
defect
Measuring range
exceeded/undercut
Cable breakage
Sensor or probe
defect
Measurement outside
permissible range
Send in for repair
Send in for repair
Stay within allowable
measurement range
No display,
unclear characters
or no response
when buttons are
pressed
Battery depleted
System error
Device is defective
Battery depleted
Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too high
Measuring cell defect
Stay within allowable
measurement range
Check the measuring cell
/ send in for repair
ERR.2
Measuring range is
undercut
Measurement too low
Measuring cell defect
Stay within allowable
measurement range
Check the measuring cell
/ send in for repair
SYS ERR
System error
Error in the device
Switch device on/off
Replace batteries
Send in for repair