Oxygen Analyzer
Version 2.11 – Revision January 2021 – Part 01381
17
YYYY-MM-DD HH:MM:SS Datalog of A–OUT
1 (O
2
*100)
Datalog of A–OUT 2 Unix time
2009-11-12
13:01:54
2089
0000
1258027314
The SD card shall be replaced after two years due to the limited capacity.
Software and default configuration uploading
The folder
bin
at the SD card includes two files:
G1.bin
and
g1parm.bin
.
G1.bin
is the
software file and
g1parm.bin
is the default configuration file.
If Green Instruments delivers a new software version with two files named
G1.bin
and
g1parm.bin,
the following steps will guide you through the installation:
Replace the old files at the SD card with the new files
G1.bin
and
g1parm.bin
that
you have received.
Set the SD card back in the analyzer. Then, restart the analyzer. Now the analyzer
will automatically check the SD card to see if there is a new software version. If yes,
the new software will automatically be loaded to the internal memory of the analyz-
er.
After uploading the new software version, reload our default settings by choosing
Load default settings in
Tools setting\Service (password level 2)\Load default set-
tings.
After uploading the new software version, you have to reset the analyzer following
the instructions in chapter 5.
3.4.7
Other Connections
The CAN connection is not in use by default.