GSI Novus™ Newborn Hearing Screening System User Manual
D-0113050 Rev H
Page 58 of 96
D
EVICE
Select
Device
to set your preference for Power Save,
Power Off, set the time and date, and to delete data
from the device.
•
Power Save (Standby) can be set to Never, 1-5
minutes or 10 minutes.
•
Power Off can be set to Never or 1, 5, 10, 15
or 30 minutes.
•
Display brightness controls the screen
brightness.
•
Set time and date allows you to set the time
and date, including the hour format (12/24),
and the date format. When complete select
the check button to return to the Device
Setup.
•
Delete all patients and test data erases all the
patient and test information from the device.
A confirmation is required before the data is
deleted.
To confirm the selection and exit the Device screen,
select the
Back
button in the bottom control bar.
S
OUND
Select
Sound
to set your preferences for the
presentation of a sound as feedback and the setting
of the volume. Conditions for which you can choose
to present a sound are:
•
Touching a button or key (function button or
keyboard key).
•
Test end – plays a sound when a test is
completed.
•
Warning – plays a sound when an electrode
becomes detached during ABR.
To confirm the selection and exit the Sound screen,
select the
Back
button in the bottom control bar.