It can be measured by converting the voltage using a shunt resistor.4-20m convert to 1-5V by using a 250Ω shunt
resistor.
Since there is a range of 1-5V, please try scaling and unit conversion with EU settings according to the sensor you
are using. * 250Ω shunt resistor (B-551-10) is sold as an option.
(SD
メモリカード
/USB
メモ
リ
)
2
.
What is the resolution of each range when measuring voltage?
Please check the table below for the resolution of each range.
3
.
What is the measurement input impedance in ohms?
The input impedance is 1MΩ.
The allowable signal source resistance is less than 300Ω."
4
.
Is it possible to input negative voltage into the analog input voltage?
This is the measurement range for analog voltage input.
In the case of the 20mV range, it is possible to measure -20mV to +20mV.
5
.
What is the resolution of the temperature?
0.1°C.
1
1. Is it possible to measure the current (4-20mA) ?
Q & A - QUESTION - midi LOGGER GL240
Frequently asked questions