. I have installed the wireless LAN unit option (B-568) and set GL240 as a station (child unit), but
GL240 is not searched even if I execute "Search" in "Select" on the GL100_240_840-APS connection
screen.
Check the following items.
・
Does the PC have a wireless network connection to the wireless LAN master device?(Please check
the network connection in the control panel.)
・
Do the SSID and security key settings on the wireless LAN master device match the settings on the
GL240?(The wireless LAN symbol on the GL240 must be enabled.)
・
When the DHCP server function is enabled on the wireless LAN master device and "Automatic IP
address acquisition" is set to "Use" on GL240, is the IP address being acquired correctly?
・
When using a static IP, is the upper segment of the PC side and the GL240 side the same?
・
When using a fixed IP, is the IP address set in the GL240 duplicated in other devices?
・
If your PC is equipped with both a wireless LAN and a wired LAN, and both are connected, disabling
the wired LAN may allow you to search correctly.
(Go to "Control Panel" → "Network and Sharing Center" → "Change Adapter Settings" → right-click on
"Local Area Connection" and select "Disable")
*If you want to use both wireless LAN and wired LAN, select "Manual Connection" from "Select".
7
.
Is it possible to send mail from GL100_240_840-APS to Gmail, yahoo mail, etc.?
It can be sent to an external server with secure functions.
However, if the GL240 is setting as an "access point", the e-mail function is not available.
8
.
What happens if there is a communication failure between the logger and the PC during recording?
The logger itself is equipped with a buffer for 1000 points.
Even if a communication error occurs at a sampling interval of x1000 points, the buffered data will
be received after communication is resumed, so no data loss will occur.
If communication is not possible for a period of time longer than the sampling interval x 1000 points,
data will be lost.
In that case, please acquire the data storage in logger itself and improve the communication
environment.