CHAPTER 4 Settings and Measurement
(6)-3 Room temperature compensation
Selection item
Description
Internal
The room temperature compensation in the GLT400 is enabled.
(Please usually select the "Internal".)
External
This is used to execute the room temperature compensation through external device.
(6)-4 Temp. Unit
Toggles the temperature unit between °C (Centigrade) and °F (Fahrenheit) for temperature settings.
When °F (Fahrenheit) is selected, calculation is performed using the following formula:
°F (Fahrenheit) = °C (Centigrade) × 1.8 + 32
Calculate the accuracy as: Centigrade accuracy × 1.8.
(6)-5 Burnout
Sets a feature which checks sensor burnout in a thermocouple.
Selection item
Description
Off
Burnout check is disabled.
On
Periodical burnout check is conducted.
During a burnout check, voltage is applied to the GLT400. Therefore, set Burnout to “Off” when GLT400 is
connected in parallel with other devices to avoid any effect from these voltages.
(6)-6 AC Line Frequency
Select the frequency of the AC line to be used.
Selection item
Description
50Hz
For an area with a power supply frequency of 50 Hz
60Hz
For an area with a power supply frequency of 60Hz
● In this setting, select a frequency for noise removal using the digital filter.
● Note that no noise in the power supply can be removed if this setting is wrong.
● For the the sampling speed where the GLT400's digital filter becomes effective, refer to “(3)-1 Sampling
interval”.