ERROR MESSAGES
A/D default: No response
from the A/D converter.
Xtrem ADPD module damaged.
Load cell input signal too high
(>20mV).
Check the scale for some load
cell / wiring default.
Load cell input signal too low
(<-20mV).
Load cell input signal out of
range (>30mV)
Load cell excitation and A/D
converter switched off
because a short-circuit.
Check wiring and load cells.
Data flash memory storage is
corrupted and cannot be reset
to the factory defaults.
Break the verification seal to
open the XTREM cover, set to
“
unlock
”
the sealing switch. The
instrument must be adjusted
and verified / calibrated again.
Calibration weight > Max.
Adjust weight should be < Max.
The division should be >10
A/D counts.
Resolution is too high. Change
the division to a higher value.
A stable measurement cannot
not be obtained for adjusting
the scale.
Increase the filter level.
Overload:
Weight > Max+9·div
Settings value cannot be
changed because the
“
sealing
switch
”
is at lock position.
Legally relevant settings cannot
be changed without breaking
the verification seal.
Negative weight (weight < -
19e)
Initial zero setting in progress
Communication error
Check baud rate setup