36
15.
FILTERING SETTINGS
Filtering options allow the scale to be configured for use in harsh
environments or for weighing moving products, such as liquid tanks or live
animals.
Possible values are from
1
up to
7
. A smaller filter level value
implies that rapid oscillations in the weight become more
visible and vice versa. Default value is level
2
.
Additional filter for weighing live animals or people that, when
moving on the load receptor, cause oscillations in the weight
indication by exerting a variable force on the load cells.
When sets to ON, the indicator
doesn’t
show intermediate
weight values when loading / unloading the scale. The result is
that the indicator retains the last stable value until the
acquisition of a new stable or zero weight.
A/D converter sampling speed. 12 sps / 50 sps (samples per
second). A higher sample rate per second will result in faster
weight acquisition but will also show short spikes due to a blow
or draft.
Output rate for sending data from the Xtrem scale to the K3X
terminal in milliseconds. Enter the value in milliseconds.
Default value is 50 ms.