AQTC8035-INS-LAB-RevB11
1
AQTC8035 TENDERCLOUD
®
STATIC AIR MATTRESS OVERLAY:
INSTALLATION & OPERATION INSTRUCTIONS
PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE
Info: The most current version of these instructions can be found online at
www.grahamfield.com
WARNING: This product is for single patient use.
WARNING: Important! Read and understand these instructions before installing or using the
Lumex Tendercloud Static Air Mattress Overlay. These instructions contain information that
will aid you in achieving maximum therapeutic results. If you do not understand any part of
these warnings, cautions or instructions, contact a healthcare professional for direction in the
use of this product. If the overlay is not properly installed, personal injury and damage to the
overlay could result.
WARNING: If components are damaged or missing, contact your dealer immediately. DO NOT
use substitute parts. Use only Lumex replacement parts. The use of non-Lumex replacement
parts could cause personal injury, property damage, and void the warranty.
WARNING: This product has a maximum weight capacity of 300 lb.
WARNING: GF Health Products, Inc. assumes no responsibility for any damage or injury caused
by improper assembly or use of this product.
ATTENTION: EMERGENCY DEFLATION PROCEDURE
Mattress must be deflated before starting CPR. For quick deflation, completely unscrew large
red valve by turning counter-clockwise.
PROCEDIMIENTO PARA DESINFLAR EN EMERGENCIA
El colchón debe de estar desinflado antes de empezar RCP (Resucitación Cardiopulmonar).
Para desinflar rápidamente, ajuste completamente la válvula roja grande volteándola hacia su
izquierda.
INTENDED USE OF THIS DEVICE
The intended use of the Tendercloud Static Air Mattress Overlay is
as a support and pressure reduction
surface.
INSTALLATION
Inflation
1. From the foot end of the bed, unroll the overlay on top of regular mattress with valve on the left side.
Smooth out and place each of the four corner straps around the standard hospital mattress.
2. Exclusive 2-part valve system:
a. Unscrew top black cap.
b. Tighten bottom red valve (it is essential that the bottom red valve be closed firmly).
3. Insert air-filling device into top of the red valve and inflate for 60 seconds if using an electric pump.
4. Replace black top cap and tighten firmly after inflation.
5. Make up bed as usual with sheets, leaving overlay valve exposed (incontinence pads may be used).
6. Patient may now be placed on overlay.
OPERATION
Hand Check
1. With patient in place (lying down), gently insert your hand palm up under overlay directly beneath the
patient’s buttocks.
2. Always keep the back of your hand flat against the hospital mattress. If the patient’s weight is directly
resting on the palm of your hand, overlay will need to be inflated for an additional 30 seconds. If you can-
not feel the patient, overlay is working efficiently and additional inflation is not required.