Operation
14
312668ZAC
Calibration
Recalibrate the meter for dispensing different fluids.
1. If the Calibration Screen is not already displayed,
complete steps 1 - 3 of the
page 12.
2. CAL
blinks indicating the Calibration Menu screen.
3. Do one of the following:
Use the current calibration.
• Press and hold the Auto /
Reset button to lock in
the displayed amount.
The display advances to
the Auto Preset Amount
Menu.
or
Recalibrate the meter:
a. If the unit of measurement is gallons, pints,
or quarts, dispense exactly one (1) quart of
fluid into a calibrated one (1) quart container.
If the unit of measurement is liters, dispense
exactly one (1) liter of fluid into a calibrated
one (1) liter container.
For proper calibration, the exact amount
must be dispensed.
b. Press and hold the
Manual / Reset button
until CAL stops blink
-
ing.
c. When CAL starts to
blink again, the display
should show 1.00,
which indicates the new calibration is com
-
plete.
NOTE: If an error was made during meter recalibration,
repeat steps a - c of the recalibration process.
d. Press and hold the
Auto / Reset button to
advance to the Auto
Preset Amount Menu.
Auto Preset Amount
Specifies an amount displayed when in Auto Dispense
Mode. Typically, enter the amount most frequently dis
-
pensed.
1. If the Auto Preset Amount screen is not already dis
-
played, complete steps 1 - 3 of the
page 12.
F
IG
. 14
GAL
QTS
PTS
LITERS
R-TOTAL
CAL