Software
English, Revision 03, Date: 20.11.2015
14
44,0,-15,9627,-3335,-256,50
44,0,-17,9626,-3333,-256,51
In this example the first number “44“ presents the current status of the device
(here: threshold 2 exceeded), then Us1= 0 and Ud1= 17, followed by Us2= 9626
and Ud2= -3333. The last displayed numbers show the collateral deviations for
antenna 1 and antenna 2. If no wire was detected, -256 is displayed as distance
value.
By using the protocol function of Hyperterm the data can be logged.
stops the
output.
-
with
the firmware can be updated, see 5.3 on page 18
-
the
ervicemenu cannot be modified by the user
5.1.2
Antenna menu
In this chapter the sub menu for antenna 1 will be explained. The sub menu for antenna
2 is identical.
Figure 5
Screenshot: Antenna menu
-
By pressing
the frequency in the range of 1 to 28 kHz can be modified.
Please
note that the antenna HG 19210-C works in a range from 3 to 25 kHz!
-
helps adjusting the threshold which refers to the sum voltage. When this
threshold is exceeded, the front LED CDx and the corresponding bits in the sys-
tem status are set.
-
By using
the distance between the guide wire and the bottom of the sensor can
be modified.
-
With
the internal height, which is specified for each type of antenna, can be
entered.
NOTE!
The sum of the values entered under
and
are used to cal-
culate the distance
S1: 1 D1: -10 S2: 10816 D2: -4403 X1:-256 mm X2: -50 mm Status: 0x40
(F)req 1 select [/Hz]: 10000
(D)etect Level 1 [/Smpl]: 1000
(H)eight of Antenna mounting [/mm]: 60
(I)nternal height [/mm]: 35
(C)alibrate
(Q)uit