GMC-I Messtechnik GmbH
89
8.6
Indication of Operating States
Control Mode / Operating State
Additional Linking
LED / Display
Color
Display Function Comment
Contin-
uous
Blinking
Voltage regulation
OUTPUT ON
CV
green
Current regulation
OUTPUT ON
CC
green
OUTPUT
ON
OUTPUT
red
OFF
OUTPUT
red
OFF
In event of OTP
OUTPUT
red
See OTP function
Primary overload, I limit
OUTPUT ON
P curve
yellow
OTP / Pon rcl
OUTPUT
OFF
P curve
yellow
Retained after manual reset
OUTPUT
red
As long as OTP is active
SRQ/STS
yellow
Retained after manual reset
OTP / Pon
OUTPUT
OFF
P curve
yellow
Retained after manual reset
stby, rst
OUTPUT
red
Remains latching OFF
SRQ/STS
yellow
Retained after manual reset
Overvoltage OVP
CV
green
OUTPUT is latching in the OFF state
OUTPUT
red
SRQ/STS
yellow
Indicates reset
Overcurrent protection OCP
selected in menu
OCP is on in the FUNCTION
menu
Iset
yellow
Brief blinking
OCP active
CC
green
Iset
yellow
Brief blinking
SRQ/STS
yellow
Indicates reset
OUTPUT
red
Remains latching OFF
Display of
Output voltage
Uout
green
Activated output voltage appears at the display
Output voltage setpoint
SET function
Uset
yellow
Uset appears at the display
Output current
Iout
green
Activated output current appears at the display
Output current setpoint
SET function
Iset
yellow
Iset appears at the display
Bus functions
Remote programming
REMOTE
yellow
Addressing mode
ADR / DATA
yellow
Service request
In remote mode
SRQ / STS
yellow
Indicates activated protective function
Front panel controls disabled
With activated function
LOCKED / SEQ
yellow
Sequence function
LOCKED / SEQ
yellow
1/1
With running sequence
LOCKED / SEQ
yellow
1/3
With suspended sequence
Measured values, menu
Two 4-place numeric LED displays