GMC-I Messtechnik GmbH
39
!
The
buS
functions menu appears.
☞
Select the
Addr
function by pressing the <
FUNCTION
> key, or
the <
FUNCTION
> and simultaneously the <
> key or the <
>
key.
!
Addr
appears at the left-hand display, and the last selected
parameter appears at the right-hand display.
☞
Select an address for the SSP KONSTANTER with the <
> or
the <
> key.
!
Any address within a range of 0 to 31 may be selected.
!
The selected address must be the same as the address
selected for the SSP KONSTANTER at the computer.
!
Address 0 is usually assigned to the controller.
!
Address 31 means unL (unlist).
☞
Acknowledge your selection with the <
ENTER
> key.
!
The display is returned to the functions menu after
acknowledging with the <
ENTER
> key.
Figure 4.7.4 a
Path to Device Address Selection
bAud – Setting Transmission Speed
•
Specifies transmission speed in bits per second.
•
The baud rate is an operating parameter of the serial interface.
•
A low baud rate should be used in the case of large distances
between communicating devices and in EMC-critical
environments.
•
Settings at the SSP KONSTANTER and the controller must be
identical.
•
The baud rate can be set to one of the following values:
50, 75, 150, 200, 300, 600, 1200, 1800, 2400, 3600, 4800,
7200, 9600 or 19,200 bits per second.
Figure 4.7.4 b
Path to Selection of the bAud Text Parameter
Settings
☞
Procedure
for selecting
bAud
and setting the desired parameter
!
bAud
appears in the
buS
display along with the related
parameter (transmission speed).
dbit – Selecting the Number of Data Bits
•
Either 7 or 8 data bits can be used.
•
Settings at the SSP KONSTANTER and the controller must be
identical.
Settings
☞
Procedure
for selecting
dbIt
and setting the desired parameter
!
dbIt
appears in the
buS
display along with the related
parameter (number of data bits).
Figure 4.7.4 c
Path to Selection of the dbIt Text Parameter
Pbit – Setting the Parity Bit
•
Parity or check bits are used to detect transmission errors.
•
They are added to the character string so that the sum of H or
L is even or odd as defined.
•
The parity bit can be set to one of the following values:
none
No parity bit
ZEro
Always 0
EVEn
Even numbered
odd
Odd numbered
nonE
Always 1
!
A parity bit should always be used in EMC-critical
environments.
!
Settings at the SSP KONSTANTER and the controller must be
identical.
Settings
☞
Procedure
for selecting
Pbit (
see also chapter 4.7
!
Pbit
appears in the
buS
display along with the related
parameter (parity bit).
Figure 4.7.4 d
Path to Selection of the PbIt Text Parameter
From the buS function group
Select the
device address
FUNCTION
Press repeatedly if necessary
CE/LOCAL
FUNCTION
ENTER
CE/LOCAL
Select the text parameter
from an options loop
From the buS function group
*
ENTER
CE/LOCAL
FUNCTION
From the buS function group
*
ENTER
CE/LOCAL
FUNCTION
Select the text parameter
from an options loop
From the buS function group
*
ENTER
CE/LOCAL
FUNCTION