74
GMC-I Messtechnik GmbH
10
Test Sequences
Default Settings
1
Depending on the setting for the “Measuring Sequences” parameter in
SETUP
2
RINS is deactivated as a default setting – can be activated via the se-
quence parameters.
3
Measurement under special test conditions deactivated as a default set-
ting – can be activated via the sequence parameters.
4
Additional testing of conductive/metallic parts which are not connected
to the protective conductor
Auto-detection
= parameter for automatic detection of the DUT’s connection (see page 78)
10.1
General
If the same sequence of single tests will be run frequently (one
after the other with subsequent report generation), for example as
specified in the standards, it’s advisable to make use of test
sequences (also called measuring sequences).
Limit values have been entered for test sequences in accordance
with the standards. And thus a go/no-go evaluation takes place
during measurement based on worst-case assessment. If the
momentary measured value is displayed in green, it lies within the
limit values specified in the standard. If the measured value is red,
is doesn’t fulfill the requirements set forth in the standard.
Note
The go/no-go evaluation of the measured values is per-
formed with greater accuracy than the value which
appears at the display, which may lead to the fact that,
due to the missing decimal places, a measured value
which appears at the display may seem to correspond
exactly to the limit value although it’s highlighted in red (as
a limit value violation) due to the places to the right of the
decimal point.
If the measured value is orange, further entries are required after
the test step (e.g. cable length), which are decisive as to whether
or not the test has been passed. Even if the DUT fails just one sin-
gle measurement, the test sequence is aborted and testing in
accordance with the selected standard is failed.
Automatic test sequences are run in rotary switch positions A1
through A9.
Test sequences A1 through A9 are preconfigured at the factory.
We recommend assigning frequently used test sequences to A1
through A9 for which parameters often need to be adjusted.
The measurements are evaluated automatically by the test instru-
ment. Evaluation is based on the worst-case and, depending on
settings, in consideration of measuring uncertainty.
Settings for the test sequences can be entered to the test instru-
ment in two different ways:
•
SETUP switch position:
general settings can be entered which
apply to all test sequences (regardless of the respectively
selected standard).
•
Switch positions A1 through A9:
classification and sequence
parameters can be entered which only apply to the selected
switch position.
Test Sequences in AUTO Rotary Switch Positions (A1 to A9)
The following test sequences are included in the test instrument’s
rotary switch positions as a standard feature:
•
DIN VDE 0701-0702
Periodic testing and testing after repair and modification of electrical
equipment
•
IEC 62353
Medical electrical equipment – Recurrent test and test after
repair of medical electrical equipment (applied parts with test
probe P1)
•
IEC 60601
(3
rd
edition)
Medical electrical equipment
– Part 1: General requirements for basic safety and essential
performance
The individual sequences are selected with the softkeys.
Automatic test sequences,
rotary switch level: orange
Switch
Setting
Standard /
Test Sequence
Measure-
ment Type
Connection
Protec-
tion
class
Freely Configurable Depending on Selected Configuration (protection
class, type of applied part)
Preconfigured (freely adjustable) test sequences
A1
IEC 62353
1
Passive
Test socket,
1 group BF APPs A-K
PC I
Short-circuit test – visual inspection – RPE
– RINS PC I – RINS APP
– RINS LN < > APP F – RINS PE < > APP F – IE PC I – IT Alt.
4)
– IA BF – function test
A2
IEC 62353
1
Passive
Test socket,
1 group BF APPs A-K
PC II
Short-circuit test – visual inspection – RINS PC II – RINS APP
– RINS LN < > APP F – RINS PE < > APP F – IE PC I – IT Alt.
4)
– IA BF – function test
A3
IEC 62353
1
Passive
Test socket,
1 group BF APPs A-K
PC I + PC II Short-circuit test – visual inspection – RPE
– RINS PC I – RINS PC II – RINS APP
– RINS LN < > APP F – RINS PE < > APP F – IE PC I – IT Alt.
4)
– IA BF – function test
A4
IEC 62353
1
Active
Auto-detection,
1 group BF APPs A-K
PC I
Short-circuit test – visual inspection – RPE – RINS PC I – RINS APP – RINS LN < > APP F
– RINS PE < > APP F – IE NL PC I – IT NL
4
– IA NL BF – IE LN PC I – IT LN
4
– IA LN BF
– function test
A5
IEC 62353
1
Active
Auto-detection,
1 group BF APPs A-K
PC II
Short-circuit test – visual inspection – RINS PC II – RINS APP – RINS LN < > APP F – RINS PE < >
APP F – IE NL PC I – IT NL
4)
– IA NL BF – IE LN PC I – IT LN
4
– IA LN BF – function test
A6
IEC 62353
1
Active
Auto-detection,
1 group BF APPs A-K
PC I + PC II Short-circuit test – visual inspection – RPE – RINS PC I – RINS PC II – RINS APP
– RINS LN < > APP F – RINS PE < > APP F – IE NL PC I – IT NL
4)
– IA NL BF – IE LN PC I
– IT LN
4
– IA LN BF – function test
A7
KA00
VDE 0701-0702
ÖVE E 8701
SNR 462638
1
Passive
Auto-detection
PC I + PC II Short-circuit test – visual inspection – RPE – RINS PC I – RINS PC II
4
– RINS APP– IPE NL
– IT NL
4
– IPE LN – IT LN
4
– function test
A8
KA00
VDE 0701-0702
ÖVE E 8701
SNR 462638
1
Active
Auto-detection
PC I + PC II Short-circuit test – visual inspection – RPE – RINS PC I – RINS PC II
4
– RINS APP– IPE NL
– IT NL
4
– IPE LN – IT LN
4
– function test
A9
KA00
VDE 0701-0702
ÖVE E 8701-EDV
SNR 462638-EDV
1
Active
Auto-detection
PC I + PC II Short-circuit test – visual inspection – RPE – IPE NL – IT NL
4
– IPE LN – IT LN
4
– function test
A7
KA01
IEC 60601
3
rd
edition
1
Active
Auto-detection,
1 group BF APPs A-K
PC I
Short-circuit test – visual inspection – RPE – RINS PC I
2
– RINS APPs
2
– IPE – IT – IP
– function test
A8
KA01
IEC 60601
3
rd
edition
1
Active
Auto-detection,
1 group BF APPs A-K
PC II
Short-circuit test – visual inspection – RPE – RINS PC I
2
– RINS PC II
2
– RINS APPs
2
– IPE
3
– IT
3
– IP
3
– function test
A9
KA01
IEC 60601
3
rd
edition
1
Active
Auto-detection,
1 group BF APPs A-E
1 group CF APPs F-K
PC I + PC II Short-circuit test – visual inspection – RPE – RINS PC I
2
– RINS PC II
2
– RINS APPs
2
– IPE
3
– IT
3
– IP
3
– function test