64
GMC-I Messtechnik GmbH
8.8
Probe Voltage – U
Mains to Test Socket
Schematic Diagram
Wiring Diagram
Permanently connected DUT
Schematic Diagram
Wiring Diagram
Direct, alternating and pulsating voltages of up to 253 V can be
measured. Two connection types are available, one of which has
to be selected in the parameters menu.
Setting Measuring Parameters for U
Probe
Test Sequence
➭
Set the rotary switch to the
U
position.
➭
Connect the DUT’s mains plug to the test instrument’s test
socket.
➭
Start the test:
press the
START
/STOP
key.
➭
PE-P1 (with mains):
Acknowledge the warning which in-
dicates that line voltage will be connected to the test
socket.
➭
Switch the device under test on.
➭
Contact the ungrounded output for safety extra-low voltage
with test probe P1.
➭
Polarity can be set via direct selection immediately be-
fore measurement is started, without having to switch to
the parameters menu.
➭
The measured values are displayed. The measured
value recording symbol shown at the right appears.
Each time this key is pressed, the currently displayed
measured value is saved to buffer memory.
➭
Turn off the device under test.
➭
End the test:
press the
START
/STOP
key.
The save symbol appears (floppy disk showing the
number of measured values stored to buffer memory)
and prompts you to save the measured values to an ID
number.
➭
Press the
ESC
key in order to discard the measured
values stored to buffer memory and acknowledge by
pressing the key shown at the right.
Single measurements,
rotary switch level: green
Switch
Po
si
ti
o
n
Measure-
ment Type
With Mains
to Test
Socket
Measure-
ment Type
Without
Mains to Test
Socket
Measuring Functions
U
PE - P1
U
Probe voltage, RMS
U
~
Alternating voltage component
U
=
Direct voltage component
PE - P1
(with mains)
U
Probe voltage, RMS
U
~
Alternating voltage component
U
=
Direct voltage component
U
Measuring
Parameter
Meaning
Measurement Type
Suitable for
DUT Connection via
PE-P1
Measurement of voltages with
reference to PE, test socket re-
mains voltage-free
Permanent connection
PE-P1 (with
mains)
Measurement of voltages with
reference to PE, line voltage is
applied to the test socket
Test socket
Polarity
Only for PE-P1 (with mains)
L/N or N/L
Selection of polarity for mains voltage to the test socket