54
Gossen Metrawatt GmbH
15.8.1 Protective Conductor Current – IPE
1
Adapter
AT3-IIIE, AT3-IIS or AT3-II S32:
Voltage measuring inputs for leakage current measurement with differential
method only with test instrument including feature I01 (e.g. SECUTEST ST
PRO)
2
Voltage measuring inputs for leakage current measurement with differential
method and use of a current clamp sensor only with test instrument includ-
ing feature I01 (e.g. SECUTEST ST PRO)
Application
Protective conductor current must be measured for protection
class I DUTs.
Definition of Protective Conductor Current (direct measurement)
Current which flows through the protective conductor in the case
of housings which are isolated from ground.
Definition of Differential Current
Sum of instantaneous current values which flow via the
L and N conductors at the DUT’s mains connection. Differential
current is practically identical to fault current in the event of a fault.
Fault current: current which is caused by an insulation defect, and
which flows via the defective point.
Definition of Alternative Measuring Method (equivalent leakage
current)
Equivalent leakage current is current which flows through the
DUT’s active conductors, which are connected to each other (L/
N), to the protective conductor (PC1) or to the exposed, conduc-
tive parts (PC2).
Differential Current Measuring Method
The device under test is operated with mains power. The sum of
the momentary values of all current which flows through all active
conductors (L/N) at the mains side of the test instrument connec-
tion is measured. The measurements must be performed with
mains plug polarity in both directions.
Alternative Measuring Method (equivalent leakage current)
A high-impedance power supply is connected between the short-
circuited mains terminals and all exposed metal parts of the hous-
ing (which are connected to each other).
Current which flows over the insulation at the device under test is
measured.
Protective Conductor Current Measuring Method (direct measure-
ment)
The device under test is operated with mains power. Current
which flows through the PE conductor to earth at the mains side
of the test instrument connection is measured.
Note
Regardless of the currently selected connection type, all
help images and schematic diagrams can be queried for
the selected measuring function.
Direct measuring method
–
Direct measurement type
– DUT mains plug to test socket
Schematic Diagram
The device under test is operated with mains power.
Protective conductor current is measured between the protective
conductor at the mains and the protective conductor terminal at
the DUT via the test instrument’s mains cable.
Wiring Diagram
Differential Current Method
–
Differential measurement type
– DUT mains plug to test socket
Schematic Diagram
The device under test is operated with mains power.
Differential current is measured between mains conductors L and
N (current clamp concept).
Single Measurements,
Rotary Switch Level: Green
Ro
ta
ry
Swi
tc
h
Po
si
ti
o
n
Meas. Type
With Mains to
Test Socket
Meas. Type
Without
Mains to Test
Socket
Measuring Functions
I
PE
Direct
I
PE
Protective conductor current,
RMS
I
PE
~
AC component
I
PE=
DC component
U
LN
Test voltage
Differential
I
PE
Protective conductor current,
RMS
U
LN
Test voltage
Alternative
I
PE
Protective conductor current,
RMS
U
Test voltage
AT3 adapter
1
I
PE
Protective conductor current,
RMS
U
LN
Test voltage
Clamp
2
I
PE
Protective conductor current,
RMS
U
LN
Test voltage
I
PE
Summary of Contents for SECULIFE ST BASE 25
Page 111: ...Gossen Metrawatt GmbH 111 ...