38
Gossen Metrawatt GmbH
13
Connecting the Device Under Test
Note
Always connect the DUT in accordance with the sche-
matic diagrams included in the online help function (see
section 7.4 “Help Functions (HELP key)”).
Connection of the DUT to the test instrument depends on:
•
The type of DUT:
For direct connection to the test socket (TS)
For DUTs with single-phase connection
and extension cords via the
EL1
adapter
(in which case the EL1 is connected to probe sockets P1)
For permanent connection (to the mains)
by contacting the housing with the probe
(for the
measurement of protective conductor resistance
or
with the direct measuring method for the touch current
measurement)
Measurement of protective conductor current
with a current clamp
(only possible with feature I01)
For connection via adapter
– With
single-phase extension cords
via the
EL1
adapter
(in which case the EL1 is connected to probe sockets P1)
– With single and
3-phase extension cords
via the
VL2E
adapter
to the test socket
– With devices with 5-pole, 16 A CEE plug
via the
AT16-DI
differential current adapter to the test socket
– With devices with 5-pole, 32 A CEE plug
via the
AT32-DI
differential current adapter to the test socket
•
DUT protection class
(PC I, PC II or PC III) or any combinations
of protection classes
Note
The DUT must be switched on for all tests. Switches,
relays, temperature regulators etc. must all be taken into
consideration.
As a default setting, the program sequence assumes that the plug
from the DUT has been connected to the test socket.
13.1
List of Possible DUT Connections
Depending on Measurement Type
Meas. Type
Suitable for DUT Connection via
RPE
PE(TS) - P1 passive
Test socket, EL1 test socket, VL2E, AT3-IIIE, AT3-IIS, AT3-
IIS32, AT16DI/AT32DI
PE(TS) - P1 active
Test socket (for PRCDs)
PE(mains) - P1
Permanent connection
PE(mains) - P1 clamp
Permanent connection
P1 - P2
Permanent connection
RINS
LN(TS) - PE(TS)
Test socket, EL1, VL2E, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/
AT32DI, CEE adapter
LN(TS) - P1
Test socket, VL2E, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/AT32DI
P1 - P2
No connection (PC3)
PE(mains) - P1
Permanent connection
PE(TS) - P1
Test socket
LN(TS) - P1//PE(TS)
Test socket, VL2E, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/AT32DI
IPE
Direct
Test socket, AT16DI/AT32DI (direct or diff.)
Differential
Test socket
Alternative
Test socket, VL2E, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/AT32DI
AT3 adapter
AT3-IIIE, AT3-IIS, AT3-IIS32
Clamp
Permanent connection
IT
Direct
Test socket, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/AT32DI
Differential
Test socket
Alternative (P1)
Test socket, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/AT32DI, VL2E
Permanent connection
Permanent connection
Alternative (P1–P2)
No connection (PC3)
IE
Direct
Test socket, AT16DI/AT32DI (only diff. is sensible)
Differential
Test socket
Alternative
Test socket, AT16DI/AT32DI
AT3 adapter
AT3-IIIE, AT3-IIS, AT3-IIS32
Clamp
Permanent connection
IA
Direct (P1)
Test socket
Alternative (P1)
Test socket
Perm. conn. (P1)
Permanent connection
IP
Direct (P1)
Test socket
Perm. conn. (P1)
Permanent connection
U probe
PE - P1
Permanent connection
PE - P1 (with mains)
Test socket
U meas.
V - COM
Permanent connection
V - COM (with mains)
Test socket
tPRCD
Mains to test socket
Test socket
P
Function test
Test socket, AT3-IIIE, AT3-IIS, AT3-IIS32, AT16DI/AT32DI, CEE
adapter
EL1
EL1 adapter
EL1 and test socket
EL1 adapter
(continuity only)
EL1 and test socket
AT3-IIIE adapter
AT3-IIIE
VL2E adapter
VL2E
Temperature
V-COM PT100(0)
Permanent connection
Current (via clamp)
V-COM
Permanent connection
V-COM (with mains)
Test socket
PRCD time to trip
—
Test socket