Gossen MetraWatt METRAHit 28C light Operating Instructions Manual Download Page 18

18

GOSSEN METRAWATT GMBH

9.1

Activating the Default Values

Previously entered changes can be undone, and the default 
settings can be reactivated. This may be advisable under the 
following circumstances:

After the occurrence of software or hardware errors

If you are under the impression that the calibrator does 
not work correctly

Simultaneously press and hold the 

ESC, 

#

 and 

"

 keys, 

and switch the instrument on with the 

ON|OFF

 key.

9.2

Transmission Mode Operation with RS 232 Interface

The calibrator is equipped with a bidirectional infrared inter-
face for communication with a PC. The commands are opti-
cally transferred through the calibrator housing by means of 
infrared light to an interface adapter (accessory), which is 
attached to the calibrator. The adapter’s RS 232 interface 
allows for the establishment of a connection to the PC via an 
interface cable. 
Commands and parameters can be transmitted from the PC 
to the calibrator. The following functions can be executed:

Configuration and read-out of calibration parameters

Calibration function and calibration range selection

Start calibration

Programming of user-specific procedures

Configuring Interface Parameters

Addr – Address

If several instruments are connected to the PC via interface 
adapters, each device must be assigned its own address. 
Address 1 should be used for the first device address, 2 for 
the second etc. If only one calibrator is used, an address 
between 1 and 14 should be selected.
Address 15 is not used for addressing, i.e. the device with 
address 15 always responds, regardless of the actual 
address.

Switching the Interface On via PC

After transmission of a data frame from the PC, the simulator 
is switched on. 
For this reason, the plug type mains supply should be used 
for long operational periods. This avoids automatic deactiva-
tion by the battery monitoring circuit.

Operating Controls in REMOTE Operation

When set to REMOTE operation, the instrument reacts like in 
local control operation. The instrument is put back into local 
control operation after switching it off and on again with the 

ON|OFF

 key.

LC Display in REMOTE Operation

The commands transmitted by the PC during direct operation 
will be executed by the instrument directly, provided the basic 
function set by the message complies with the one set on the 
rotary switch. The LC display will acknowledge the program-
med signal value and the output type after each message.
Incorrect settings of the rotary switch are signalled by a flas-
hing display of the scheduled basic function. The new com-
mand will be executed immediately when the rotary switch 
has been set to the scheduled basic function.

Interface parameters are as follows: 
Format:

8 data bits, no parity, 1 stop bit

Baud rate:

9600 baud

Signal LED on the Interface Adapter

One green and one red signal diode on the interface adapter 
give you the following information and warnings:

Green

 LED:

Lights up briefly when commands are transmitted from 
the PC to the calibrator.

Red

 LED:

Lights up during the transmission of messages from the 
calibrator to the PC. This occurs each time the calibrator 
is switched on and when a message received from the 
PC is acknowledged. This response, as well as an 
acousting signal, are generated to confirm the correct 
physical communication between PC and calibrator.

Summary of Contents for METRAHit 28C light

Page 1: ...Operating Instructions METRA HIT 28c light Calibrator 3 349 317 15 6 2 05 ...

Page 2: ...nnector jack for NA5 600 power pack 8 Connector jacks for calibrator output 9 DKD calibration mark see page 4 Symbols used in the Digital Display 1 Main display with decimal point and polarity display 2 Auxiliary display with decimal point and polarity display 3 Simulator in continuous operation 4 NUM Numeric entry of output signal 5 INT Interval sequence active 6 RAMP Ramp function active 7 ON Ca...

Page 3: ...Fuses 24 12 4 Housing 24 13 Calibrator Messages 24 14 Repair and Replacement Parts Service DKD Calibration Lab and Rental Instrument Service 25 15 Warranty 25 16 Product Support 25 1 Safety Features and Precautions You have selected an instrument which provides you with a high level of safety This instrument fulfills the requirements of the applicable European and national EC guidelines We confirm...

Page 4: ...t back to the cali brator In order to avoid damage to the instrument when interfer ence voltages are applied within allowable limit values the mA SINK and mA SOURCE measuring circuit is equipped with a fuse F3 which makes this measuring circuit highly resistive if excessive current should occur in the event of a fault for the duration of overloading Warning The instrument may not be operated in ex...

Page 5: ...is shown in the diagram on page 3 The instrument is ready for calibration as soon as the key is released Switching the Instrument On with a PC After transmission of a data frame from the PC the simulator is switched on See also chapter 9 2 Note Electrical discharge and high frequency interference may cause incorrect displays to appear and may dis able the simulator In such cases switch the instru ...

Page 6: ...nt of one another with the frequency generator The output signal is a square wave The resistance of the interconnected circuit should not be any less than 1 kΩ Select the Hz calibration function with the rotary switch Setting voltage amplitude 0 15 V Press the ESC key until you proceed to the menu for set ting voltage amplitude The decade i e the position of the digit to be changed is selected wit...

Page 7: ... the position of the digit to be changed is selected with the keys and the respective digit is set with the keys Note The following error messages may appear HiCurr high current current too high where I 4 5 mA and LoCurr low current current too low or polarity reversal where I 40 µA i e open jack sockets Attention In this operating mode interference voltages must not be applied to the calibrator j...

Page 8: ...tern Press the key to return to the temperature si mulation value main menu The measured internal refer nece temperature is indicated with an I for internal on the right hand side at the bottom of the auxiliary display extern Press the key to proceed to the temperature value sub menu Setting the external temperature value The last entry position on the right hand side blinks Enter the digits with ...

Page 9: ...of a high interference voltage the in strument calibrator is protected by a replaceable fuse F3 see chapter 12 3 Example ad a Internal Reference Junction Example ad b External Reference Junction Selecting Sensor Type Setting the Temperature Simulation Value and Entering the External Reference Temperature Device to be calibrated Terminal Block Equalizing Lead CALIBRATOR Output Device to be calibrat...

Page 10: ...t sink operating mode because thermal overload would otherwise occur and the fuse F3 would blow LoVoltappears at the display where the voltage is too low Example of a 2 Wire Transmitter Measuring Circuit 7 2 Current Source Internal supply power is used for the simulation of a current source Note The internal control circuit of the current source is monitored if voltage drops by more than 20 V at t...

Page 11: ...and recorders during single handed operation Input parameters for interval sequences include All simulator functions except for Hz can be adjusted as output quantities A lower Start and an upper End range limit can be selected for each output quantity from within the overall range The number of steps can be set within a range of 1 99 9 The number of steps can be entered as a whole number as well w...

Page 12: ...nu Flowchart on page 12 2 The sequence is stopped by pressing the or the key Interval time elapsed thus far is saved as tx 3 The sequence is resumed by pressing the key and remaining sequence duration ty is equal to t1 tx 4 Stop the interval sequence by simultaneously pressing and holding the and the key 2 audible acoustic sig nals must be generated Menu Flowchart Starting and Stopping the Interva...

Page 13: ... or temp 0 20 Int range mA Range 0 20 4 20 0 24 10 000 Int start mA 20 000 Int end mA 05 5 Int steps Number of 02 00 Int t1 auto Int no Interval Duration Periodic Sequence YES no Start End Output Quantity SEt must blink Configure the ramp Lower Range Limit Upper Range Limit 2x Interval Steps t1 s ESC ESC ESC ESC ESC ESC ESC ESC ...

Page 14: ...ences Once only t1 t2 t3 Repetitive t1 t2 t3 t2 t1 t2 t3 Examples of Ramp Sequences Example of a Periodic Ramp Sequence Manually Controlled Ramp Sequences After configuring all parameters in accordance with the menu flowchart on page 15 rising or declining ramps can be trig gered with the and the keys The relationship between the output signal and each of the key operations is depicted with the he...

Page 15: ...mp 0 20 ramp range mA Range 0 20 4 20 0 24 10 000 ramp start mA 20 000 ramp end mA 00 05 ramp t1 Rise Time t1 02 00 ramp t2 00 28 ramp t3 repeat ramp yes Dwell Time t2 Periodic Ramp YES no Start End Output Quantity SEt must blink Configure ramp Lower Range Limit Upper Range Limit 2x ESC ESC ESC ESC ESC ESC ESC ESC Decline Time t3 ESC ...

Page 16: ... in order to change the cor responding parameter or parameters After the desired digit has been selected with the keys and the value has been adjusted with the keys the next digit is accessed with the key or the display is returned to the start menu or switched to the next sub menu The calibration mode is started by repeatedly pressing the ESC key until the simulator display appears The calibrator...

Page 17: ...tIme 15 set addr ESC Change Value Set Device Address Acknowledge Address 15 50 05 set tIme ESC Change Value Set Time Acknowledge Entry 20 11 01 set date Set Date Select Position Possible Settings TT MM JJ TT day MM month JJ year Possible Settings hh mm ss hh hour mm minute ss second Digit blinks Possible Setting Values see also chapter 9 2 0 15 set Change Value Select Position ESC set ESC Acknowle...

Page 18: ...ardless of the actual address Switching the Interface On via PC After transmission of a data frame from the PC the simulator is switched on For this reason the plug type mains supply should be used for long operational periods This avoids automatic deactiva tion by the battery monitoring circuit Operating Controls in REMOTE Operation When set to REMOTE operation the instrument reacts like in local...

Page 19: ...less documentation and man agement of calibration results the creation of calibration pro cedures and remote control of the calibrator METRA HIT 28c light sequence controls can be imple mented online or off line after downloading complete calibra tion procedures The following conditions must be fulfilled in order to allow for use of METRAwin 90 2 Hardware Requirements WINDOWS and IBM compatible PC...

Page 20: ...laces With Load of Intrinsic Error Over load Direct Voltage Simulator of s mV Imax V 0 300mV 0 01 mV 700 Ω 0 05 0 02 18 mA 2 0 3 V 0 1 mV 1000 Ω 0 05 0 2 0 10 V 1 mV 1000 Ω 0 05 2 0 15 V 1 mV 1000 Ω 0 05 2 Pulse Frequency Generator Keying ratio mark to space ratio 50 amplitude 10 mV 15 V of s Hz Imax Hz 1 Hz 1 kHz 0 1 8Hz1 1000 Ω 0 05 0 2 18 mA Current Source Max Load Impedance of s µA mA 4 20 mA ...

Page 21: ...reased by a factor of 1 8 The F sub ranges are calculated as follows F 32 C x 1 8 Additional Error for Thermocouple Simulation Thermocouple Type T Error in K for Thermocouple Types at Ref Temp 0 C Sub Range C J L T U K E S R B N 200 100 1 17 0 83 1 52 1 2 1 59 1 03 2 38 100 0 0 55 0 56 0 78 0 77 0 73 0 51 50 4 79 50 5 29 1 03 0 100 0 42 0 41 0 52 0 51 0 53 0 35 3 77 3 92 0 77 100 200 0 46 0 45 0 4...

Page 22: ...ination Factor 2 Test Voltage 500 V per EN 61010 1 2001 VDE 0411 1 2002 Electromagnetic Compatibility EMC Interference emission EN 61326 2002 class B Interference immunity EN 61326 2002 IEC 61000 4 2 Feature A 8 kV atmospheric discharge 4 kV contact discharge IEC 61000 4 3 Feature B 3 V m Data Interface Data Transmission optical via infrared light through the housing with interface adapter as acce...

Page 23: ...ordance with IEC R 6 or IEC LR 6 or equivalent rechargeable batter ies Replacing the Batteries Set the instrument face down onto a flat working surface loosen the two screws at the back and lift off the housing base starting at the bottom The housing top and hous ing bottom are held together with the help of snap hooks at the top front Remove the batteries from the battery compartment Insert three...

Page 24: ...for each series METRA HIT 28c light calibrator If a unit of measure blinks this indicates that the balancing constant which has been established and saved to the calibrator is no longer available for the respective function If this is the case results may deviate from the specification We recommend sending the instrument to our Repair and Replacement Parts department for rebalancing see chapter 14...

Page 25: ... expertise in the field of metrol ogy from test reports and proprietary calibration certificates right on up to DKD calibration certificates Our spectrum of offerings is rounded out with free test equip ment management An on site DKD calibration station is an integral part of our ser vice department If errors are discovered during calibration our specialized personnel are capable of completing rep...

Page 26: ...change without notice A pdf version is available on the Internet GOSSEN METRAWATT GMBH Thomas Mann Str 16 20 90471 Nürnberg Germany Phone 49 0 911 8602 0 Fax 49 0 911 8602 669 E Mail info gossenmetrawatt com www gossenmetrawatt com ...

Reviews: