13.19 Measurements with Test Adapter
1
Differential current method only with test instruments
including feature I01 (e.g. SECUTEST ST PRO)
2
Alternatively AT3-IIS or AT3-II S32 for IPE and IE
Attention!
!
Please read and observe the operating
instructions for the test adapters regard-
ing correct connection of the test
adapter and the DUT, as well as pecu-
liarities involved in the test procedure.
Connection Example with EL1
Connection Example with VL2E
Connection Example for Measuring Protective Con-
ductor Current IPE
1
(direct method) with AT3-IIIE
Connection Example for Measuring Device Leak-
age Current IG
1
(differential current method) with
AT3-IIIE
1
Only with feature I01(e.g. SECUTEST ST PRO or SECULIFE ST BASE(25))
Test with Adapter
EL1
VL2E AT3-
IIIE
2
AT16DI
AT32DI
CEE
adapter
Connectors for the DUT
Non-heating devices,
1P+N+PE 16 A
—
✔
✔
—
—
Earthing contact,
1P+N+ PE 16 A
—
✔
—
—
—
CEE, 1P+N+PE 16 A
—
✔
✔
—
✔
CEE, 3P+N+PE 16 A
—
✔
✔
✔
/ —
✔
CEE, 3P+N+PE 32 A
—
✔
✔
— /
✔
✔
5 x 4mm sockets
—
—
—
—
✔
Connectors for the test instrument
Earthing contact,
1P+N+ PE 16 A
—
—
✔
✔
—
Socket for test probe
—
✔
✔
—
—
Plug for V–COM
1
—
—
✔
—
—
Active test
Protective conductor current
IPE
– Direct method
—
—
✔
✔
—
– Differential current
—
—
✔
1
✔
—
Device leakage current
IE
– Direct method
—
—
✔
✔
—
– Differential current
—
—
✔
1
✔
—
Touch current
T
—
—
✔
✔
—
Passive test
Pro. con. resistance
RPE
✔
✔
✔
✔
✔
Insulation resistance
RINS
✔
✔
✔
✔
✔
Protective conductor cur-
rent
IPE
(equivalent leak-
age current method)
—
✔
✔
✔
✔
Extension cords:
the following measurements, in addition to RPE
& RINS, are performed in switch position
EL1
Single-phase (3-pole)
✔
✔
✔
—
—
3-phase (5-pole)
—
✔
✔
—
—
Wire short-circuit
✔
✔
✔
—
—
Wire break
✔
✔
✔
—
—
Reversed wires
—
✔
✔
—
—
EL1 Adapter
P1
V
COM
P1