SECU
TEST ST
B
ASE
(
10
) /
P
RO
and SECU
LIFE ST
BASE
(25)
14
Gossen Metrawatt GmbH
Safety
shutdown
At DUT differential current of
> 10 mA,
shutdown time: < 500 ms,
can also be set to > 30 mA,
in the event of probe current
(electronic fuse) during:
– Leakage current measure-
ment: > 30 mA~/< 500 ms*
– Protective conductor resis-
tance measurement:
> 250 mA~/< 1 ms
in case of continuous current
I > 16.5 A
Fuse
links
Mains fuses:
2 ea. FF 500V/16A
Special fuse:
M 250V/250mA
10 A RPE test current
(feature G01) only:
1 ea. FF 500V/16A
Electromagnetic Compatibility
Product standard EN 61326-1
EN 61326-2-2
USB Data Port
Type
USB slave for PC connection /
remote control**
Type
2 ea. USB master,
for data entry devices * with HID boot
interface,
for USB flash drive for data backup,
for USB flash drive for saving
reports as HTML files, for printers *
* See operating instructions for compatible test in-
struments
** Remote control only with extension: “Remote
Control via PC (IZYTRONIQ)” (included as stan-
dard feature with SECUTEST ST PRO and avail-
able with SECUTEST DB+ (Z853R or feature
KB01).
Bluetooth
®
data interface 2.1 + EDR
(test instruments with feature M01 only)
Frequency range Max. 2.5 mW (class II)
Transmission
intensity
2400 … 2483.5 MHz
Mechanical Design
Display
4.3" multi-display
(9.7 × 5.5 cm), backlit,
480 x 272 pixels with 24 bit
color depth (true color)
Dimensions
W × H × D:
295 ×145 × 150 mm
Height with handle: 170 mm
Weight
SECUTEST ST BASE(10)/
PRO: approx. 2.5 kg
SECULIFE ST BASE25:
Approx. 4.0 kg
(depending on the test instru-
ment version)
Protection
Housing: IP 40,
Test socket: IP 20
per EN 60529
Database
Number of
data records
50,000
1 data record = 1 DUT or
location node or cus-
tomer or individual mea-
surement
6.1
Relevant Standards
The test instrument has been manufactured
and tested in accordance with the following
safety regulations:
*
Firmware version 3.2.0 and lower: 12 mA
EN 61010-1 Safety requirements for electrical equipment
for measurement, control and laboratory use
– Part 1: General requirements
EN 60529
Test instruments and test procedures
Degrees of protection provided by enclo-
sures (IP code)
EN 61326-1 Electrical equipment for measurement, con-
trol and laboratory use – EMC requirements
– Part 1: General requirements
EN 61326-2 Electrical equipment for measurement, con-
trol and laboratory use – EMC requirements
– Part 2-2: Particular requirements – Test
configurations, operational conditions and
performance criteria for sensitive test and
measurement equipment for EMC unpro-
tected applications
EN 61557-16 Electrical safety in low voltage distribution
systems up to 1000 V AC and 1500 V DC –
Equipment for testing, measuring or moni-
toring of protective measures –
Part 16: Devices for testing the effectiveness
of protective measures of electrical devices
and/or electrical medical devices