37
Gossen Metrawatt GmbH
METRACELL BT PRO
10.5
Charge Measurement
With the help of the
CHARGE
measuring function, the voltages of the blocks of a battery can be measured several times at short intervals
while charging (capacity test of the blocks).
Note
At least two test runs must be completed for the
charge
measurement.
Note
Polarity can be switched off
28. If polarity is switched off, the absolute value of the measured value is recorded.
Se-
quence
Description of the Measurement
Display
1
Connect the multimeter test probe measurement cables to the P–/COM and S+ sockets with
the matching colors (
2
Press and hold the
FUNCTION
key for about 2 seconds in order to switch the battery tester on.
BTpro
<date>
<version>
<time>
3
Press any key to open the
main menu
.
If you’re already in another menu, press the
FUNCTION
key repeatedly until the
main menu
ap-
pears.
->
MEASUREMENT
DATA
4
Press the
OK
key in order to open the
MEASUREMENTS
menu.
The
MEASUREMENTS
menu is displayed.
->
MULTIMETER
FLOAT
5
Select
CHARGE
using the
▼ ▲
keys.
DISCHARGE
->
CHARGE
6
Press the
OK
key in order to open the selected menu for the measuring function.
->
START
SELECT.BATT
7
Before starting the measurement, the location, battery and limit value parameters must be set.
These can be:
–
Selected in the battery database via RFID
step 7a.
–
Selected manually in the battery database
step 7b.
–
Entered manually at the battery tester
step 7c.
START
->
SELECT.BATT
7a
A battery database has been created
The battery database has been transferred to the battery tester
You use RFID technology for your system.
The DUTs in the battery database are linked to the corresponding RFID tags
1. Select
SELECT.BATT
with the
▼
key.
2. Press the
OK
key
The battery database is displayed. An acoustic signal indicates that the RFID reader is
ready.
3. Read the RFID tag (
An acoustic signal confirms reading and the DUT is displayed.
4. Press the
OK
key.
The battery is selected. The battery tester returns to the
MEASUREMENTS
menu.
START
->
SELECT.BATT
7b
A battery database has been created
The battery database has been transferred to the battery tester
1. Select
SELECT.BATT
with the
▼
key.
2. Press the
OK
key
The battery database is displayed. An acoustic signal indicates that the RFID reader is
ready.
3. Select a battery using the
▼ ▲
keys.
4. Press the
OK
key.
The battery is selected and the battery tester returns to the
MEASUREMENTS
menu.
START
->
SELECT.BATT
7c
1. Select the
AREA:XXXX.
parameter with the
▼
key.
2. Press the
OK
key.
The entry mode is active.
3. Enter the parameter’s first digit
4. Acknowledge your entry with the
OK
key.
5. Repeat steps 1 through 4 for all other parameter digits for battery number, number of
blocks and upper and lower voltage limit values.
SELECT.BATT
->
AREA
:
0
000
8
Return to
START
using the
▲
key.
->
START
SELECT.BATT
9
Press the
OK
key in order to start the measurement.
BLOCK
- - - - - - -
XXXX/X
AN:XXXX
BN:XXXX