REVIEW Screen (Just Go)
11.
Review the performance measurements with the patient. You have the option of entering
Notes
about the task on this screen.
12. Select one of the following options:
◉
back To PaTIeNT seTUP:
Return to the
PaTIeNT seTUP
screen.
◉
assIGN To eXIsTING PaTIeNT:
Open the
Save Data to Patient Profile
screen in order to log
the collected performance measurements to an existing patient.
◉
reTUrN To JUsT Go:
Return to the
Get Patient ready
screen.
Recover from a Prevented Fall
The purpose of this procedure is to reset the system after a prevented fall.
When a patient fall is prevented by the system, the
falls Prevented
screen displays and the actuator LED
status light bar changes to either blinking yellow if fall detection was due to exceeding the DFP Sensitivity
setting, or to blinking green if fall detection was due to exceeding the Descent Limit. The actuator does
not allow any further downward motion. In order to allow the patient to correct their position, the system
maintains tension on the strap while allowing for upward movement and horizontal motion at a slower
speed. During a task, the user can choose whether or not to record the prevented fall into the system.
This option is not available during
JUsT Go
.
NoTIce:
Ensure that the patient has completely regained his or her balance before
clearing the Falls Prevented alert.