PosITIoNING aNd HorIzoNTal lock
37
Contact Gorbel Medical
| Phone: 1-844-846-8744 | E-mail: [email protected]
Position the Spreader Bar
The spreader bar should be positioned low enough to attach the patient harness via the spreader bar
straps, without coming into contact with the patient’s head (1”-2”). The therapist should hold the spreader
bar securely to prevent contact with the patient’s head.
Depending on where you accessed the
PosITIoNING
screen from, Select a Patient or press
beGIN
sessIoN, eXIT sessIoN
or
JUsT Go
to continue.
1. If necessary, touch
PosITIoNING.
2.
From the
spreader bar Position
area of the
PosITIoNING
screen, touch and hold the up or down arrows in order to
move the spreader bar to the desired position.
3. After connecting the patient harness to the spreader bar,
tighten the spreader bar straps until there is tension. Ensure
that the straps are symmetrical and that the spreader bar is
parallel with the floor.
NoTe:
an error alert will appear if the amount of upward
movement applied during this action creates and uplifting
force on the patient.