I S A – A M S O p e r a t i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 35 / 76
www.go-sys.de
9.2.2.7 Fingerprint (new functions)
from including AMS software version 4.3.4.0 and BlueBox firmware version 4.02.14
In previous software versions the fingerprint is an additionally stored absorbance spectrum.
Now the fingerprint is the enveloped area between the curves of the maximum values and the minimum val-
ues of the extinction spectra recorded after opening the spectra window.
In the absorbance spectra view there is a new Button:
Click on this Button opens and closes the Min-Max spectral values view.
The minimum and maximum values of the single wavelengths of the absorbance spectra recorded
after opening the spectra window are displayed.
Red line: maximum values curve
Green line: minimum values curve
Enveloped area (marked grey here)
Number of absorbance spectra recorded after opening the spectra window
You can offset the maximum values curve and the
minimum values curve at specific wavelength
ranges. Right mouseclick in the spectra view field
opens an offset selection menu.
In [Set high offset] and set [Set low offset] you can
determine a wavelength range and an offset value.
If you are in zoom mode, the wavelength range will
shift accordingly.