Page 1: ......

Page 2: ...2 ...

Page 3: ...Y NEED Devices for radio frequency RF Beauty 6000 are produced and distributed by Domino S r l Via Vittorio Veneto 52 31013 Codognè Treviso ITALY Tel 0438 7933 Fax 0438 793363 info globuscorporation com www globuscorporation com Please refer to the guide for the RE version rechargeable WEB RECHARGE Rev 01 17 13 01 2017 ...

Page 4: ...D CONTRAINDICATIONS 19 Mandatory behavior 19 Warnings before use 19 Warnings during the use 20 Side effects 21 Contraindications 21 EQUIPMENT 22 CONNECTIONS 23 Device 23 Accessories 23 DEVICE USE 24 Switching on off the device 24 Main menu 25 User menu 25 Program menu 27 The treatment areas are the following 27 Special 27 Calibration 28 New program menu 30 Free program menu 32 My programs menu 33 ...

Page 5: ... AESTHETIC PROGRAM LIST 37 Benefits in the aesthetic field 51 Mechanism of energy transfer 58 How to apply the handpiece 59 MAINTENANCE AND CLEANING 70 Device and accessories 70 Transportation 71 FAQ 72 ...

Page 6: ...HM150PS48C2 8 PRI 100 240Vac 50 60Hz 1 8A SEC 48Vdc 3 13A Weight 600 g Supplied accessories Neutral electrode or plate Steel plate with secure connection Size 150 x 110 mm weight 216 g Monopolar resistive handpiece Ergonomic handpiece for resistive treatments The handpiece is designed for a plug in connection which facilitates the resistive electrode insertion Size 110 x 70 mm weight 230 g ...

Page 7: ...isolated only on the upper surface so as to protect the contact between the operator s hand and the plate The measures available for both electrodes are Ø 70 50 30 mm Bipolar handpiece The bipolar handpiece allows treatment with resistive mode without the need to use the neutral plate as the counter electrode The conductive surface is composed of two contact points in biocompatible nickel the inne...

Page 8: ...m Adapter that connects to the cable A two cables B with Ø 4 mm connection The adapter is useful when you want to use more than two automatic electrodes Resistive and neutral Cable B for connection between the adapter and two fixed electrodes with Ø 4 mm connection Cable C which serves as adapter allowing the connection between a 4 mm female connection and a 2 mm male connection Resistive and neut...

Page 9: ...ct Dimensions 80x800 mm Soft bag for storing cables and electrodes Active neutral handpiece kit The active neutral handpiece is an accessory that allows to use the neutral counter electrode in a dynamic way By using this accessory you can make a treatment gripping the neutral handpiece with one hand and the capacitive or resistive handpiece with the other In this way it will be possible to obtain ...

Page 10: ...ck to the retailer when purchasing a new device Warranty conditions The device is guaranteed to the first purchaser for 24 months from purchase date against defects in material or workmanship 12 months if the user uses the device for professional aims provided that it is used properly and maintained under normal operating conditions Warranty coverage is limited in the following cases six 6 months ...

Page 11: ...arged 5 The warranty becomes void if the damage has been caused by impacts falls erroneous or improper use of the product use of non original external charger or power supply accidental events alteration replacement detachment of the warranty seals and or tampering with the product The warranty does not cover damages caused during the transportation when unsuitable packages are used see point 1 6 ...

Page 12: ...nvironments other than domestic ones and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes The estimated useful life of the product is of 5 years We recommend returning the device to the manufacturer and or authorized center every two years for the maintenance and to verify the safety The device can be used continuously ...

Page 13: ... household waste management company or the store where the product was purchased It indicates that the product has been produced respecting the directive 2011 65 EEC It indicates the optimal temperatures for the storage and transportation of the product It informs the operator that before using the device he she must read the manual It indicates a compulsory procedure Alarms disabled Ionizing emis...

Page 14: ... and its accessories are stored and transported Device Input PWR AHM150PS48C2 8 PRI 100 240Vac 50 60Hz 1 8A SEC 48Vdc 3 13A PIN 1 2 V Output 48Vd c 2 2A 105VA PIN 3 4 GND Frequency Output 400 470 700 1000KHz Max Output Power 120W 10 330ohm 470KHz Type RF6K RF6K311500408 SN RF6K311500408 ...

Page 15: ... Total time 20 00 F1 total F2 total Head 50 mm Remaining time 20 00 Current Frequency 860 mA 740 kH Panel and keyboard Display and Interface Cellulite Default user program lifespan remaining time user emission power emission graph current sound alert off emission frequency transfer mode contextual menu short circuited alert off emission handpiece emission percentage ...

Page 16: ...a B si C do D re E mi F fa G sol CDE alarm it sounds when the device is emitting and it repeats every 9 seconds CDE EF open or short handpiece alarm it is related to visual alarms but it is visualized after 2 Meaning Icons for signaling contact between the handpieces non contact between the handpieces and short circuited handpieces During treatment it will be warned an alarm sound at 20 second int...

Page 17: ...wo electrodes are in contact Alerts off It is possible to turn the short circuited and sound alerts off If these alarms are disabled their icons will be displayed during the execution of the programs During the execution of the treatment on small areas and with uneven surfaces for example the hand ankle the device can record an excessive production of current triggering a protection alarm If it ne...

Page 18: ...ditions Internal temperature of 60 C indicated by a fixed orange icon The power is limited to 65 in respect to the set point in normal conditions Internal temperature of 65 C indicated by a fixed red icon The power is limited to 50 in respect to the set point in normal conditions Internal temperature of 70 C The power is automatically lowered to 0 and an acoustic signal CDE FG CDE FG is repeated e...

Page 19: ...es in force in your country Make sure that the power supply will be in a comfortable position and that it will be easy to be removed Before disconnecting one of the electrodes pause the device or turn the supply to 0 Do not modify the device Before performing the treatment on a table containing metal parts ensure that it is connected to the ground and has a proper capacity Warnings before use Dang...

Page 20: ...o its surface it must have no scratches that can cause burns Before starting the treatment the operator must be sure that the subject is capable of indicating his sensitivity to the heat The subject must take off all the bracelets necklaces piercings or metal objects The subject must take off his her acoustic prosthesis and mobile dental prosthesis if near the area to be treated Before starting th...

Page 21: ...00 on subjects with pacemakers or other electronic devices Do not treat cephalic area abdomen gonads and sexual organs Do not treat the eyes zone Do not use the device on pregnant subjects Do not treat in correspondence of resorbable and non resorbable fillers Do not use the device on body parts which are not sensitive to the heat or on subjects suffering from one of the following diseases neoplas...

Page 22: ... or plate G Resistive capacitive electrodes H Monopolar resistive handpiece I Bipolar handpiece User s manual and warranty When opening the package please check that the basic equipment is complete If some elements should be missing contact immediately your authorized retailer Control the integrity of the device and its accessories carefully All the provided information can be modified without pre...

Page 23: ...plug it as shown below First connect the connector to the device and then to the mains WARNING insert the connector with the rectangular part downwards Accessories Counter electrode connection plate Connect the counter electrode to the black cable through the plug with the red end Connect the black plug to the output D OK NO A B C D ...

Page 24: ...nserting the heads in the respective handpiece make sure that the power level is set to 0 Optional accessories Bipolar handpiece connection Bipolar handpiece ends with a black and a blue jack which have to be plugged in the outputs D and C For the connection of the accessories that make up the kit for the automatic treatment refer to the instructions contained within the kit itself DEVICE USE Swit...

Page 25: ...is section permits to manage the users Programs program list with preset protocols New program it creates new programs Free program shortcut to start a treatment in a very quick manner My programs user created programs are stored Contextual menu to access the settings of Rf Beauty 6000 User menu The user menu allows you to individually manage the device even when it is shared among several co work...

Page 26: ...t deletes a user Select Press the arrows to select the name and then press OK to confirm The selected user will appear in the upper right corner To return to the previous menu press the white arrow back New user Use the keypad to add a new user After typing the name push the enter button in the lower right corner Default user User selection Mark Default user Keyboard ...

Page 27: ...cutaneous ageing especially for eyes and smile wrinkles Skin treatments for body wrinkles and stretch marks acne scars or furuncles Cellulite treatments to heal different stages of cellulite Special The area contains some preset programs Scan the program consists of three phases in resistive mode each programmed with different frequencies The goal is to convey uniformly the energy in the most supe...

Page 28: ...ou have the need to facilitate the absorption of active ingredients of a medication The modulation in small pulses determines a temporary increase in the permeability of the tissue membranes and allows a better penetration of substances through the skin Automatic The program contained in this area should be used to carry out the treatments in automatic mode using the special kit purchased optional...

Page 29: ...f 2 Handpiece not in contact Do not hold the handpiece and plate in contact between them or with the subject Press Start Phase 1 2 At this point the device will take a few seconds to complete the control of this first phase and then switches to the next phase Phase 2 of 2 Handpiece in contact Apply the plate and handpiece to the subject press the button Start Phase 2 2 and perform movements of the...

Page 30: ... mistake for example the head choice and you want to repeat the process When the calibration is completed the impedance reading and communication between the device and the electrodes will be optimized according to their size New program menu Selecting the New program area it will be possible to create customized programs Calibration completed Confirm Cancel Calibration How many phases does the pr...

Page 31: ...program At the beginning you will be asked to choose the number of phases of the program The choice can be made by choosing a maximum of 3 phases For each of the phases you can choose whether to use the capacitive or resistive mode and set the frequency and the phase duration Capacitive F1 Phase treatment type Resistive F1 Phase duration F3 F2 20 min F2 F2 ...

Page 32: ...the calibration The program will be stored in the device memory and will be available in the My programs menu Frequency choice Rf Beauty 6000 offers four different emission frequencies In addition to traditional 470 KHz there are other available emissions 400 kHz 700 kHz and 1000 kHz A higher frequency acts primarily on surface tissues while lower frequencies can penetrate better Free program menu...

Page 33: ... treated tissues The trend graph of this phenomenon is represented by the blue curve If the program has more phases it will be possible to skip to the next phase by pressing a few seconds the right button of the keyboard Power setting The power can be increased or decreased by using the arrows on the control handle As an alternative option use the four arrows on the lower part of the keypad to inc...

Page 34: ...o pause the treatment press the specific key on the keypad pause button Every time that the treatment will be suspended the power will be reset to zero If you choose to resume treatment you will have to increase the power To resume the program push Resume To return to the main menu press Home To return to the calibration menu press the Back button Treatment in pause mode Resume Home ...

Page 35: ...e language of the device Volume Adjustment you can adjust the volume of alarms by choosing among the four proposed levels Recharge this section refers only to the Rf Beauty 6000 RE device rechargeable Alarms it can turn the sound signals on or off Disable alarms If the user during program execution wants to disable the sound alerts he she can select the contextual menu and proceed as indicated Tre...

Page 36: ... 00 F1 total F2 total Head 50 mm Remaining time 20 00 If you confirm your choice the software will automatically return to the working screen and you will notice the disabled alarm symbol on the upper right corner Disable alarm Confirm Cancel Disable Alarm short circuited handpiece Cellulite ...

Page 37: ...ood heat sensation At the end of the treatment the face must appear reddened PRECAUTIONS Do not pass the handpiece on the eyeball and eyelids The skin should be clean and cleansed EYE WRINKLES HANDPIECE MEDIUM INTENSITY DURATION PROTOCOL Bipolar 4 5 6 7 for each side of the face 2 sessions with an interval of 24 hours between them Repeat after 21 days Apply a sufficient amount of conductive cream ...

Page 38: ...clean and cleansed FACE LIFTING HANDPIECE HIGH INTENSITY DURATION PROTOCOL Bipolar 7 8 9 12 for each side of the face 2 sessions with an interval of 24 hours between them Repeat after 21 days Apply a sufficient amount of conductive cream on the face and on the handpiece Massage the skin in a continuous motion which starts from the inside of the face and goes outward Treat the forehead around the e...

Page 39: ...btain a lukewarm temperature PRECAUTIONS Do not pass the handpiece on the eyeball and eyelids The skin should be clean and cleansed COUPEROSIS HANDPIECE LOW INTENSITY DURATION PROTOCOL Bipolar 1 2 3 12 for each side of the face 2 3 times a week for 3 weeks Apply a sufficient amount of conductive cream on the face and on the handpiece Massage the skin affected by couperosis in a circular motion Adj...

Page 40: ... of the treatment the face must appear reddened PRECAUTIONS Do not pass the handpiece on the eyeball and eyelids The skin should be clean and cleansed SMOOTH MASSAGE HANDPIECE LOW INTENSITY DURATION PROTOCOL Big monopolar capacitive 1 2 3 12 for each area 2 3 times a week for 3 4 weeks The massage program helps to reduce muscle tension improve superficial microcirculation and cellular exchanges Pl...

Page 41: ...n You must perceive a moderate heat sensation The program lasts 6 minutes for each area For larger areas repeat the program BODY WRINKLES HANDPIECE HIGH INTENSITY DURATION PROTOCOL Medium monopolar capacitive 7 8 9 7 for each area 2 sessions with an interval of 24 hours between them Repeat after 21 days The program is indicated to treat body wrinkles Place the plate near the area to be treated For...

Page 42: ...of conductive cream on the electrodes and make motions above the stretch marks The program lasts 10 minutes for each area CLEAVAGE LIFTING HANDPIECE HIGH INTENSITY DURATION PROTOCOL Medium monopolar resistive 7 8 9 8 2 sessions with an interval of 24 hours between them Repeat after 21 days The cleavage with the aging process gradually tends to become wrinkled and lose elasticity This program helps...

Page 43: ...ear and zigzagging motions on the areas to be treated The program lasts 10 minutes for each area Repeat the program as needed The feeling of heat must be very strong FIRMING HANDPIECE HIGH INTENSITY DURATION PROTOCOL Big monopolar capacitive 7 8 9 8 for each area 2 3 times a week for 3 4 weeks This program is indicated to raise the tissues and tone the skin that presents some laxities Place the pl...

Page 44: ...ITE HANDPIECE LOW INTENSITY DURATION PROTOCOL Big monopolar capacitive 1 2 3 15 for each area 2 3 times a week for 6 8 weeks Big monopolar resistive 1 2 3 10 for each area The fibrous cellulite is the second stage of cellulite the skin is tougher the orange peel effect is present even without pinching the skin Place the plate near the area to be treated Apply a sufficient amount of conductive crea...

Page 45: ...ece and make the treatment massaging the skin in a continuous and circular motion You must perceive a good heat sensation At the end of 12 minutes the program can be repeated on another area for example before a buttock and then the other one BODY PIMPLES HANDPIECE LOW INTENSITY DURATION PROTOCOL Medium monopolar capacitive 1 2 3 6 for each area 2 3 times a week for 3 4 weeks This program is indic...

Page 46: ...good heat sensation At the end of 12 minutes the program can be repeated on another area for example before a buttock and then the other one MALE FEMALE ABDOMINAL ADIPOSITY HANDPIECE MEDIUM INTENSITY DURATION PROTOCOL Medium monopolar capacitive 4 5 6 10 for each area 2 3 times a week for 6 8 weeks Big monopolar resistive 4 5 6 10 for each area This program is suitable for treating the fat that ac...

Page 47: ... treatment must be repeated on another area for example before a buttock and then the other one SCLEROTIC CELLULITE HANDPIECE LOW INTENSITY DURATION PROTOCOL Big monopolar capacitive 1 2 3 17 for each area 2 3 times a week for 6 8 weeks Big monopolar resistive 1 2 3 8 for each area The sclerotic cellulite is the third stage of cellulite the skin has hollows and nodules the fibrous septa thicken Pl...

Page 48: ...f two phases to be performed in succession At the end of the two phases the treatment must be repeated on another area for example before a buttock and then the other one LYMPHATIC DRAINAGE HANDPIECE INTENSITY DURATION PROTOCOL Big monopolar capacitive 4 5 6 12 for each area 2 3 times a week for 6 8 weeks This program is indicated to improve the blood flow and drain liquid accumulations Place the ...

Page 49: ...orange peel skin such as legs gluteal muscles arms abdomen Place the plate near the area to be treated Apply a sufficient amount of conductive cream both on the plate and on the handpiece and make the treatment massaging the skin in a continuous and circular motion You must perceive a moderate heat The program is composed of two phases to be performed in succession At the end of 12 minutes the tre...

Page 50: ...umulates electrical charges the current is progressively reduced to zero when the system is charged Now if the capacitor polarity is reversed electrical current will be generated This reversed current will charge the capacitor with opposite polarity A continuous alternating polarity in the generator leads the electrical charges to shift from a plate to the other At a biological level the capacitor...

Page 51: ...olecules begin to rotate with the same frequency of the field to which they are subjected This rapid rotation occurs in the opposite direction to the electrical field and it generates a molecular friction and therefore heat Benefits in the aesthetic field In the aesthetic field instead the main benefits of radio frequency allow to improve the tone and elasticity of the skin stimulate the productio...

Page 52: ... is reflected on the function of elastin which is the protein that gives elasticity to the tissues Elastin in fact can be stretched and stiffened due to ageing process We have seen that the radio frequency results in a local temperature increase in the irradiated area Treatment with temperatures of 40 42 C can provide an initial contractions and shrinkage of collagen fibers Indeed the structure of...

Page 53: ...phic and healthy the complexion more natural there is a reduction in the number and depth of wrinkles and overall the entire face takes on a more refreshed appearance Neck and cleavage the RF used on the neck brings considerable improvements on the reduction of wrinkles The precaution is to avoid the thyroid area Abdominal laxity as a result of weight loss the abdomen skin is relaxed as in the pos...

Page 54: ...ss common in males From an aesthetic point of view cellulite is characterized by the modification of the surface of the skin that presents hollows and the classic orange peel or mattress structure The most affected areas by cellulite are hips gluteus thighs knees and the lower abdominal area With cellulite the connective tissue and localized fat undergo some alterations resulting from changes in t...

Page 55: ...men can find relief only upwards while the network shape of the male fibrous septa allows the fat to expand into other directions Other studies also show that the fat lobules space delimited by the fibrous septa in women are broader and wider than those of man The appearance of cellulite can be divided into four phases Initially there is an alteration of the precapillary sphincters rings of smooth...

Page 56: ...ation of macronodules causing sclerosis hardening of tissue From a clinical point of view cellulite can be divided into 4 stages Stage 0 there are no changes in the skin Stage 1 the skin of the affected area is smooth when the subject is standing or lying but some small waves of the skin can be seen pinching the skin or during a muscle contraction Stage 2 the skin appears in a mattress or orange p...

Page 57: ...at advantage of these devices is not to be invasive indeed to be very comfortable Several authors agree that the heat caused by the devices of radio frequency causes various physiological benefits such as the formation of new collagen a lifting effect a decrease in localized fat a reduction of edema and fibrosis and ultimately an improvement of cellulite We have already seen that the collagen subj...

Page 58: ...with a non insulated metal head The electrode surfaces are spread with cream to allow a uniform electrical conduction The alternating electrical field causes the movement and concentration of electrical charges within the tissues With the resistive mode this phenomenon produces a temperature increase especially in correspondence of the tissue with a greater impedance generating first of all a ther...

Page 59: ...ce neck and cleavage Capacitive mode The capacitive mode is done using the monopolar capacitive handpiece and the counter electrode The mobile electrode is coated by a layer of insulating material This enables the electro chemical reaction and the thermal effect to develop near the insulated electrode which acts on the underlying tissues Therefore the capacitive mode is more indicated to stimulate...

Page 60: ...eld whose direction is reversed up to a million times per second The frequency is a very important parameter in the treatments of radio frequency as when the frequency varies the depth of action also changes By using a low frequency the treatment is able to penetrate tissues and to act in depth by using instead a higher frequency the treatment will work more on the surface These indications are of...

Page 61: ...ove from one pole to the other and the effect of the temperature increase does not would be focused on our goal the thigh fat Thus especially in the resistive mode it is essential that the positioning of the neutral plate is done in order to take into account the objective of the treatment Modulation The device has some programs with a modulated emission where the machine does not emit continuousl...

Page 62: ...different between the modulated and continuous emission This phenomenon is amplified as the t off increases Pulsed Rf Beauty 6000 has the following pulsed preset programs Pulsed 90 t on 900 ms t off 100 ms Emitted power and average power Average power Emitted power Graphic representation of a pulsed emission Graphic representation of a continuous emission ...

Page 63: ...he stage at which you are acting as well as the geometry of the electrodes used All preset pulsed programs can be used in resistive mode Athermy The athermy program has a duty cycle that allows you to work safely without incurring a thermal effect In a fraction of the time t off is four times higher than the t on and this allows tissues to have the time to dissipate the excess energy The athermy p...

Page 64: ...ted with the sensation of warmth perceived and it is therefore essential to be able to distinguish at least three levels of intensity for determining the best treatment To monitor the subject during treatment and get feedback on the perceived temperature we recommend that you use a temperature scale with values ranging from 0 to 10 where zero means there is no feeling of heat while 10 corresponds ...

Page 65: ...65 Rev 01 17 11 01 2017 1 NO HEAT SENSATION 2 BARELY PERCEIVED HEAT 3 LIGHT HEAT 4 LUKEWARM 5 VERY LUKEWARM 6 HOT 7 VERY HOT 8 TOO MUCH HEAT 9 EXTREMELY HOT 10 SEARING ...

Page 66: ...case is lower than 38 C The effect of the medium power is primarily to stimulate the process of lipolysis The temperature measured on the surface of the dermis is around 38 39 C The effect of the high power is mainly to increase the voltage re consolidation of the collagen and the production of new fibers Another important effect of the high power is to stimulate the release of specific HSP molecu...

Page 67: ...ined as the time of the actual treatment once reached the recommended feeling of warmth For this reason we suggest preparing the area to be treated by heating it a few minutes with the free program in the capacitive mode and then switch to run the out and out program Treatment area for aesthetic programs Our preset programs provide a scheduled duration according to the handpiece used and the area ...

Page 68: ...68 Rev 01 17 11 01 2017 21 cm Big monopolar electrode 70 mm Ø 14 cm ...

Page 69: ...69 Rev 01 17 11 01 2017 Medium monopolar electrode 50 mm Ø 9 cm 6 cm 15 cm 10 cm Small monopolar and bipolar electrode 30 mm Ø ...

Page 70: ...device nor open it Only specialized and authorized centers can repair it Avoid violent impacts that may damage the device and lead to malfunctioning even though it may be not immediately detectable Use this device in a dry open air environment Do not wrap the device Clean the device and accessories only with disinfectant or with quaternary ammonium salt diluted with distilled water equal to 0 2 0 ...

Page 71: ...very use or at least once per week It is recommended that you perform the maintenance of the device and its parts every 12 months Place the device and its accessories in their original package when they are not used or have to be transported Transportation At the end of the treatment or during the transportation place the device and its accessories in the original carrying case ...

Page 72: ...ature that I have to transfer to the subject It is important to note that the increase of the temperature should be very low We recommend that you use the temperature scale During treatment often ask the subject which is the perceived heat sensation on a scale from 1 to 10 it is advisable not to exceed the value 5 6 If the subject responds 7 8 lower the power so that it reaches the ideal value Sho...

Page 73: ...part must be replaced Is it advisable a massage after treatment Certainly a hand massage at the end of the treatment is recommended but not required The massage has the function to relax the muscle tissue favoring the drainage of liquids How to choose the electrode size to be used In the aesthetic field the small electrode is indicated for areas such as the face or localized treatments such as sca...

Page 74: ...solutely not recommended to put the handpiece in contact with the reference plate Can the thickness of the epidermis affect the perception of heat Of course If your skin has a reduced thickness you will have a greater perception of heat ...

Page 75: ...75 Rev 01 17 11 01 2017 ...

Page 76: ...76 Rev 01 17 11 01 2017 ...