background image

6" All Stainless

 

Electronic Digital Caliper

OWNER’S MANUAL

FOR YOUR SAFETY AND CARE OF THIS TOOL,

 

please read these instructions carefully and retain them for future use.

CARE AND MAINTENANCE

1

.  Keep body face clean, prevent liquid material from getting into slider to destroy electronics.

2. 

Face should be cleaned with a clean, dry, lint-free cloth. Lubricate body with a few drops of 
clock oil. Acetone and alcohol must not be used.

3. 

Keep battery compartment clean and free of corrosion.

TROUBLESHOOTING

PROBLEM/

Every second 5 digits

jump simultaneously

Display cannot count

Display shows 000.00 

or IN 00.000

Function buttons not active

Reading error for full 

length is ≥ 0.1mm

No display on LCD screen

SOLUTION/

Battery voltage is under 1.45V. 

Remove cover and replace battery.

Faulty circuit. Remove battery, after 30 sec. 

put battery back into compartment.

Function buttons and sliders signal end may have 

shorted out. Remove the cover, put the button springs in 

place with rubber covers. Contact must be unobstructed.

Springs or rubber covers may be out of shape due to 

excessive pressing. See solution above for tips.

Sensor may have dirt or deposits in it. Remove cover 

and slider assembly. Blow off sensor face with clean 

pressurized air ≥ 51kg/cm squared), clean with a dry, 

lint-free cloth.

1. Poor battery contact. 

Check battery compartment, clean if necessary.

2. Battery voltage is under 1.3V. 

Replace with correct battery.