![Gima Rescue LIFE User Manual Download Page 58](http://html1.mh-extra.com/html/gima/rescue-life/rescue-life_user-manual_2228086058.webp)
Distributed by
Rescue Life - User Manual Rev. 3.1
52
USING A CUFF THAT IS THE WRONG SIZE MAY GIVE FALSE AND MISLEADING RESULTS
.
THE
NIBP
MODULE
IS
DESIGNED
TO
WORK
WITH
SUNTECH
CUFFS
AND
HOSES.
THE
USE
OF
CUFFS
AND
HOSES
NOT
SUPPLIED
BY
SUNTECH
MAY
COMPROMISE
PERFORMANCE
AND
ACCURACY.
D
O NOT
ATTACH THE
CUFF TO A LIMB
BEING USED FOR
IV
INFUSIONS AS THE CUFF INFLATION CAN BLOCK
THE INFUSION
,
POTENTIALLY CAUSING HARM TO THE PATIENT
.
If the blood pressure cuff is on the same limb as a pulse oximeter probe, the oxygen
saturation results will be altered when the cuff occludes the brachial artery.
Intra-arm differences vary between people. Do not assume that measurements from
both arms are same.
When a cuff is going to be positioned on a patient for an extended length of time, be
sure to occasionally check the limb for proper circulation.
ACTIVATION AND OPERATION OF THE NIBP MODE
To activate the NIBP
mode, press the ‘MENU’ (F5) key form the operational Rescue Life
screen.
On the menu select NIBP than ON using the jog.
WHEN IN NIBP MODE
,
IT IS POSSIBLE TO SELECT ENERGY AND DEFIBRILLATE
.
IF THE DEFIBRILLATOR IS
MEASURING THE PRESSURE WHEN ENERGY IS CHARGED PRESSURE MEASURING WILL BE STOPPED
TO RETURN TO THE DEFIBRILLATOR FUNCTIONS AND SETUPS PRESS
‘
EXIT
’
(F5)
KEY
.
On the right side of the graphic screen the following NIBP values are displayed:
Patient type:
‘ADULT’ (default), ‘PEDIATRIC’ or ‘NEONATAL’
Systolic pressure:
‘SYS’
Diastolic pressure:
‘DIA’
Mean arterial pressure:
‘MEAN’
Cuff pressure:
‘PRESS’
Operation mode:
‘MANUAL’
(default),
‘AUTO’