Cardiogima 1M         QUICK GUIDE             

Rev. Feb  2008                          

Cardiogima 1M         QUICK GUIDE          

Rev. Feb 2008

 

GIMA SPA                                   

www.gimaitaly.com

                                                                   GIMA SPA                                   

www.gimaitaly.com

 

 
              
             (1) Paper Feed 
 
             (2) Sequential Mode or Printing Memory Data 
 
             (3) Manual Mode 
 
             (4) Isochronous (Automatic) Mode 
 
             
             (5) Paper Speed  
 
             (6) Sensitivity of Recording  
 
             (7) Anti Muscle Tremors Filter 
 
 

(8) Anti Base Line Disturbs Filter  

 

(9) Anti Power Line Disturbs Filter 

 
 

(10/11) Vertical Arrovs: Traces Group Selection   

 

 

 

 

              Modes Menu Selection 

 

 

 

 

 

 

 

 

 

 

 

 

(12) Modes Menu 

 

 
Power on/off: 

 

To  turn  on/off  the  instrument  press  the  main  switch,  placed  on  its  back  side.  The 
initials screen will display the battery status bar. 
If the charge value is lower than 80%  please plug in the charger. 
To start the work press any key. 

 
 

Default operational set-up: 
 

At power on the default set-up is: 

• 

Traces print number:  3 

• 

Traces Order: Standard 

• 

Paper speed:  25mm/s 

• 

Sensitivity of recording:  5mm/mV 

• 

Filters:  Disabled 

• 

Hardcopy frame length: 5 seconds 

 
 

The main  operational modes: 
 

1. 

Paper  speed: 

25mm/s  at  the  turning  on.  Pushing  the  key 

(5)

 

it  is  possible  to 

select the paper speed: 50mm/s, 5mm/s,  10 mm/s,25mm/s…..5mm/s……

 

The mode appears on the display.

 

 

2. 

Sensitivity of recording:

 10mm/mV by default. Pushing the key 

(6) 

it is possible 

to select the sensitivity of the trace: 20mm/mV, 5mm/mV,…..10mm/mV….

 

The mode appears on the display. 
 

3. 

Filters:

 

disabled by default. Pushing the keys (7); (8); (9); it is possible to enable 

the desired filter:

 

 

LP

: anti muscle tremors filter 

HP

: anti base line disturbs filter 

N5/N6

: anti power line disturbs filter (N5 for 50 Hz and N6 for 60 Hz) 

 

4. 

Traces  Group: 

the  Standard  modality  is  predefined  at  power  on  and  the  display 

showes the I trace. 

 

It is possible to select the desired trace pushing the keys (10) and (11)

  .

   

With the 

Standard Modality

 the display will show: I;  II;  III; aVR;  aVL;  aVF; 

  V

1

;  V

2

;  V

3

; V

4

;  V

5;  

V

6

 

Changing to 

Cabrera Modality

 the display will show

aVL;  I;  -aVR;  II;  aVF;  III;    

    V

1

;  V

2

;  V

3

; V

4

;  V

5

;  V

6

 

 

How to open the thermal 
printer 
 
Lift the lever, acting as indicated 
by the arrow, until the cover 
group is released from its locking 
position. 
To avoid damages to the lever do 
not use excessive force 