1 Safety 

1.1 Instructions for safe operations   

 

Check  the  main  unit  and  all  accessories  periodically  to  make  sure  that  there  is  no  visible 

damage  that  may  affect  patient’s  safety  and  monitoring  performance  about  cables  and 

transducers. It is recommended that the device should be inspected once a week at least. When 

there is obvious damage, stop using the monitor. 

 

Necessary maintenance must be performed by qualified service engineers ONLY. Users are not 

permitted to maintain it by themselves. 

 

The Electrocardiograph cannot be used together with devices not specified in User’s Manual. 

 

When the cardiac defibrillator be used with other electric stimulate devices, Please select the 

electrode (type: A30-G26-21) provided by our company. 

 

The electrode and its connector of this ECG must keep from other conductor, for example, the 

earth. 

 

According the Standard IEC60601, this device is belonged to internal electrical power source, 

CF applied part. it is applied for cardiac examination. This device is safe and reliable. 

 

Please  select  the  ECG  cable  which  is  provided  by  our  company,  if  the  defibrillation  time  is 

over  5  s,  you  should  use  one-off  chest  electrode,  which  can  avoid  the  metal  electrode  burn 

patient’s skin. Please don’t use it with electric stimulator at the same time, if it is compulsory, a 

professional technician must guide on the scene. 

1.2 Warnings 

 

Explosive hazard—DO NOT use the Electrocardiograph in environment with inflammable gas 

such as some ignitable anesthetic agents. 

 

When  other  device  be  joined  with  this  ECG  device,  it  must  be  accorded  with  Standard 

IEC60601,  internal  electrical  power  source.  because  creepage  summation  may  injure  the 

patient, the creepage current be inspected by the other device. 

 

The disposal of scrap instrument and its accessories (including battery) should follow the local 

laws and regulations. 

1.3 Attentions 

 

Keep  the  Electrocardiograph  away  from  dust,  vibration,  corrosive  substances,  explosive 

materials, high temperature and moisture. 

 

If the Electrocardiograph gets wet, please stop operating it. 

 

When it is carried from cold environment to warm and humid environment, please do not use it 

immediately. 

 

Do not operate keys on front panel with sharp materials. 

 

Do not have the product immerged in liquid. 

 

When cleaning the device with water, the temperature should be lower than 60

Summary of Contents for 33257

Page 1: ...Code 33257 CARDIOPOCKET ECG USER MANUAL ...

Page 2: ... for details Please read the Manual very carefully before using this equipment These instructions describe the operating procedures to be followed strictly Failure to follow these instructions can cause measuring abnormality equipment damage and personal injury The manufacturer is NOT responsible for the safety reliability and performance issues and any monitoring abnormality personal injury and e...

Page 3: ...rview 2 3 Principle and Caution 3 4 Technical specifications 4 5 Accessories 4 6 Installation 5 7 Operating Guide 8 8 The notice of the battery operating 12 9 Repairing and Maintenance 13 10 Troubleshooting 14 11 Key of Symbols 17 ...

Page 4: ...ded by our company if the defibrillation time is over 5 s you should use one off chest electrode which can avoid the metal electrode burn patient s skin Please don t use it with electric stimulator at the same time if it is compulsory a professional technician must guide on the scene 1 2 Warnings Explosive hazard DO NOT use the Electrocardiograph in environment with inflammable gas such as some ig...

Page 5: ...0 05 150Hz frequency response which is the world level also the equipment has advanced digital baseline drifting filter Alternating Current filter and Myo electricity disturbance filter The sampling rate of lead signal is 800Hz the closure frequency of low pass filter machine with guards against mixes folds is 3dB 220Hz The sampling rate is higher than the standard of not lower than 500Hz which ma...

Page 6: ...tem coordinates the operation of print controlling equipment liquid crystal screen and keyboard driver it is responsible for the profile print driver controlling information demonstration keyboard management and some other control of CMS80 3 2 Caution 3 2 1 Observe the status of the patients and the instrument momentarily 3 2 2 Use the patient cable only to connect the patient and instrument 3 2 3...

Page 7: ... provided by the factory It may damage this equipment if using the accessories with different models Power adapter Input AC 100 240V 50 60Hz Output DC 9V Attention User can choose the power cord according the request of local power net and also the safe standard Power supply adapter should be accorded with relevant regulation according to IEC60601 1 One lithium rechargeable battery DC 7 4V lithium...

Page 8: ... 5 6 Installation 6 1 Sketch on each side and accessory Front view ...

Page 9: ...d V4 V4 Fifth inter costal space at left mid clavicular line V5 Left anterior axillary line at the horizontal lever of V4 V6 Left mid axillary line at the horizontal lever of V4 Clean the positions around to which chest electrodes are to be attached with alcohol lay the ECG medical gel on dia 25cm area around the position and the edge of chest electrode cap press the suction bulb firmly attach the...

Page 10: ...d skin Attention Tighten the knob after inserting the mainframe patient cable pin in the patient cable socket 6 2 3 Electrode connection definitions and color code Electrode Location Electrode Code Line Color Electrode Color Socket Number Right arm R Black Red 9 Left arm L Black Yellow 10 Left leg F Black Green 11 Right leg RF Black Black 14 Chest 1 V1 C1 White Red 12 Chest 2 V2 C2 White Yellow 1 ...

Page 11: ...ing paper according to the illustration then clip them in the relevant position in paper fixer 3 Close the paper cabinet cover and protrudes 2cm in the top of the recording paper 7 Operating Guide 7 1 The instrument starts self checking after its turning on In the self checking process 7 2 The instrument will show the following state after self checking under DEMO mode ...

Page 12: ...ning for example then press the button to switch the relevant information this column is read only The state and other indicator information are all read only Mode Conversion shows the mode conversion of the systemic lead it divides into Auto lead conversion AUTO and Manual lead conversion MANU Leading condition indication it will show NORM when the leads collecting signals normally and OVER when ...

Page 13: ...tton as followings Press button to return after installing paper 5 Calibrating Press button to print a lmv standard voltage mark in order to observe the current sensitivity directly The LCD will display the following picture when in calibrating process In calibrating process Please Waiting English Interface Ending the Calibrating English Interface ATTENTION The calibrating process is operated auto...

Page 14: ...trument records the twelve patient cables automatically three seconds ECG signal each patient cable Filter OFF 50Hz 60Hz 50Hz 60Hz totally five types of filter among them 50Hz and 60Hz denote the opening of 35Hz EMG filter ATTENTION Switching on the EMG interference filter will fall the range of the recorded R wave Speed 25mm s 50mm s totally two types of paper skipping speed mode System Manual En...

Page 15: ...When the battery is full the device can be operated for one hour when the battery be used as power supply An icon of battery will be displayed in the LCD screen of front panel this icon indicates power of battery When the battery is power off the device will turn off automatically this setting is for avoiding permanent damage on battery caused by excessive discharge 3 please charge the battery aft...

Page 16: ...t guarantees for one year and do not arbitrary disassembly Remove the electric plug when there is power cut the instrument must be kept cool and dry when it is unused for long time electrify the instrument every three months Inspect and maintain of Lead electrode 1 The state of leads can be checked by multimeter please check the leads according to following list If the resistance between electrode...

Page 17: ...eck battery capacity The instrument has the function of protecting the electro circuit when the battery over discharged 2 Check the AC power voltage The instrument has the function of protecting the electro circuit when its overvoltage 3 Check AC interference check the knob on the plug of the patient cable The instrument has the function of protecting the electro circuit when over loading 10 2 AC ...

Page 18: ... electrodes and patients body 4 Keep the patient from motion or hyperventilation 5 Check the connection between patient cable and electrodes Please use filter if still having above mentioned interference 10 5 Troubleshooting Trouble Possible Reason Solution Big interference Random waveform 1 It is firm whether the grounding connections 2 The joint of two sides of conductance line is not stable 3 A...

Page 19: ...nd leader s plug part is not connection enough 5 the connection between patient cable and electrodes is not well 1 The alcohol of use is asked for well 2 Use ethyl alcohol to wash electrode wafer and the skin under the electrode 3 Be charged to the battery 4 The electrode reed tightly clip Baseline drift 1 The batteries are drained 2 Patient moves 1 Be charged to the battery 2 Keep patients quietl...

Page 20: ...n key power supply acknowledgement shortcut menu Function Key Pause On Function Key System Menu Function Key Print Function Key Calibrate AC power is on no battery The battery solely powering and the battery is full of electricity The battery solely powering and the battery is not full of electricity The battery solely powering and the electric power is going to be used up On charging Danger volta...

Page 21: ...rts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components subject to wear are not included in the warranty The repair or replacement performed during the warranty period shall not extend the warranty The warranty is void in the following cases repairs performed by unauthorized personnel or with non original spare parts defects caused by ne...