Technical Specifications
Please be aware of the following before operating the instrument.
PLC Purification System with Integrated ELS Detector
SPeCifiCation
Definition
Communication
USB (internal)
Control
GGP software installed on the integrated touchscreen PC
Data rate
Up to 30 Hz
Detection
Photodiode with high sensitivity
Filter
Moving average
Gain setting
Dynamic gain management: SAGA (SEDEX Automated Gain Adjustment)
Gas requirements*
*For oil-free, dry and filtered
gas, nitrogen, or air.
SPECIFICATION
DEFINITION
Pressure
2–4.5 bar (29–67 psi)
Purity
Gas purity has negligible impact on ELS detector performance
Tubing
Polyamide (PA), 6 mm (OD)
Fitting
Stainless steel (SS), 6 mm with push-lock fitting
Light source
Selected high efficiency blue LED (470 mm) with elapsed-time counter
Nebulizer*
*Model SEDEX FP™ (Flash/
Prep)
SPECIFICATION
DEFINITION
Flow rate range
100 µL/min to 5 mL/min
Temperature range
Ambient to 100°C
Weight*
*In addition to a standard
PLC Purification System
5 kg (11 lbs.)