SETTING THE VOLUME

 

MyPIPETMAN®

Chapter 2

SETTING THE VOLUME

The volume of liquid to be aspirated is set using the volume display. The dials are colored either black or 

red to indicate the position of the decimal point, depending on the model (refer to Figure 4 below).

Volume selection differs according to your configuration. To obtain maximum accuracy when setting the 

volume, proceed as follows:

 

When 

decreasing

 the volume setting, slowly reach the required setting, making sure not to 

overshoot the volume mark.

 

When 

increasing

 the volume setting, pass the required value by 1/3 of a turn and slowly decrease to 

reach the volume, making sure not to overshoot the volume mark.

MODEL

COLOR OF VOLUMETER NUMBERS

BLACK

RED

INCREMENT

MyPIPETMAN SINGLE CHANNEL 

MyPIPETMAN P2

µL

0.01 µL

0.002 µL

MyPIPETMAN P10

µL

0.1 µL

0.02 µL

MyPIPETMAN P20

µL

0.1 µL

0.02 µL

MyPIPETMAN P100

µL

-

0.2 µL

MyPIPETMAN P200

µL

-

0.2 µL

MyPIPETMAN P1000

0.01 mL mL

0.002 mL

MyPIPETMAN P5000

0.01 mL mL

0.002 mL

MyPIPETMAN P10mL

mL

0.1 mL

0.02 mL

MyPIPETMAN MULTICHANNEL 

MyPIPETMAN P8x10

µL

0.1 µL

0.02 µL

MyPIPETMAN P12x10

µL

0.1 µL

0.02 µL

MyPIPETMAN P8x200

µL

-

0.2 µL

MyPIPETMAN P12x200

µL

-

0.2 µL

MyPIPETMAN P8x300

µL

-

0.2 µL

MyPIPETMAN P12x300

µL

-

0.2 µL

1

5

 

P2

1.25 µL

2

0

7

5

 

P10

7.5 µL

1

2

5

 

P20

12.5 µL

 

P100

0

7

5

75 µL

 

P200

1

2

5

125 µL

 

P1000

0

7

5

0.75 mL

 

P5000

1

2

5

1.25 mL

 

P10mL 

0

7

5

7.5 mL

0

7

5

 

x10

7.5 µL

1

2

5

 

x20

12.5 µL

 

x200

1

2

5

125 µL

 

x300

1

2

5

125 µL

Multichannel 

model

Single channel 

model

Figure 4

Dial colors by model

Simple Ratchet

For pipettes mounted with a simple ratchet, select the volume 

by turning the push button (refer to Figure 5).

3-Position Locking System

For pipettes equipped with a 3-position locking system, turn the 

selector to your desired position (refer to Figure 6).

Free volume selection position

1. 

Turn the selector to the left until the cursor reaches the 
unlocking 

1

.

2.  Select the volume by turning the push button.

Light click volume selection position

1. 

Turn the selector until the cursor reaches the middle 

2

.

2.  Select the volume by turning the push button.

Figure 5

Simple ratchet

Figure 6

3-position locking 

system

NOTE

When selecting the volume, you will hear 

click sounds, allowing for more precise 

settings

.

 Use these sounds for 

final fine 

volume selection.

3

2

1

Secure Pipetting

After selecting your volume, turn the selector to the right 
until the cursor reaches the locking 

3

 for secure pipetting.

NOTE

To ensure good performance, always lock 

the pipette before pipetting by selecting the 

locking icon.   

 

Summary of Contents for MyPipetman

Page 1: ...User s Guide EN A U TOCLAVABL E FULLY...

Page 2: ...r Good Pipetting 10 ACCESSORIES 11 MyPIPETMAN Accessories for Extra Personalization 11 TROUBLESHOOTING 14 Troubleshooting Table 14 Leak Test 15 CLEANING AND DECONTAMINATION 15 Cleaning for Single Chan...

Page 3: ...tte Colors Special Text Printing Special Logo Printing Ordering Information Minimum Order Quantity Available Countries 1 Worldwide Possible on Request Subject to Specific Conditions 150 Units Mix of D...

Page 4: ...and MyPIPETMAN Enterprise serial number type A A 50001 MyPIPETMAN serial number type MY 524 A A Figure 2 Serial number identification example Year Month Production number Model Random Number Month Ye...

Page 5: ...6 0 3 MyPIPETMAN P200 D200 20 100 200 0 50 0 80 1 60 0 20 0 25 0 30 2 5 0 8 0 8 1 0 0 25 0 15 1 6 1 6 1 6 0 6 0 6 0 6 8 0 1 6 0 8 3 0 0 6 0 3 MyPIPETMAN P1000 D1000 100 500 1000 3 0 4 0 8 0 0 6 1 0 1...

Page 6: ...00 L 0 2 L MyPIPETMAN P8x300 L 0 2 L MyPIPETMAN P12x300 L 0 2 L 1 5 P2 1 25 L 2 0 7 5 P10 7 5 L 1 2 5 P20 12 5 L P100 0 7 5 75 L P200 1 2 5 125 L P1000 0 7 5 0 75 mL P5000 1 2 5 1 25 mL P10mL 0 7 5 7...

Page 7: ...r extension refer to Figure 7 1 Slide the extension over the tip holder 2 Push the extension firmly onto the end of the tip ejector until it clicks into place To remove a tip ejector extension 1 Gentl...

Page 8: ...accuracy and precision by creating an efficient barrier to aerosols or corrosive vapors such as acids which could damage mechanical parts Insert the filter directly in the pipette tip holder Change th...

Page 9: ...wall of the recipient vessel at an angle of 10 to 40 2 Press the push button slowly and smoothly to the first stop 3 Wait for at least a second then press the push button to the second stop to expel...

Page 10: ...the tip For volatile solvents saturate the air cushion in your pipette by aspirating and dispensing the solvent repeatedly before aspirating the sample When the liquid temperature is different from t...

Page 11: ...nap cap and screw cap microtubes Description Part Number JIMMY Microtube Opener set of 3 F144983 For P5000 and P10mL models disposable filters maintain the pipette s original accuracy and precision by...

Page 12: ...P200 MyPIPETMAN FP070142 Metal Ejector Dark Blue P1000 MyPIPETMAN FP070151 Metal Ejector White P1000 MyPIPETMAN FP070152 PLASTIC TIP EJECTORS Model Part Number Plastic Ejector Dark Blue P2 P10 MYPIPET...

Page 13: ...Holder Dark Blue P1000 MyPIPETMAN FP070281 Tip Hholder White P1000 MyPIPETMAN FP070282 Tip Holder Dark Blue P5000 MyPIPETMAN FP070291 Tip Holder White P5000 MyPIPETMAN FP070292 Tip Holder Dark Blue P...

Page 14: ...place the tip holder S Connecting nut is loose Tighten connecting nut S M Damaged or corroded piston Return pipette to supplier S Improper repair or assembly Refer to MAINTENANCE S Pipetted volume is...

Page 15: ...pipette to the maximum volume given in the specifications and pre wet the tips 3 Aspirate the set volume from a reagent reservoir of distilled water 4 Maintain the pipette in the vertical position and...

Page 16: ...or wipe them with a soft cloth or lint free tissue soaked with the cleaning solution 3 Rinse the components with distilled water 4 Wipe the entire pipette with a soft cloth or lint free tissue soaked...

Page 17: ...e being immersed in decontamination solution 1 Disassemble the pipette refer to MAINTENANCE 2 Immerse tip ejector tip holder and connecting nut in the decontamination solution 3 Disassemble and degrea...

Page 18: ...p ejector along the tip holder 3 Clip the tip ejector on the body of the pipette Changing the Tip Holder for Single Channel 1 Push the ejection button 2 Turn the tip ejector clockwise 3 Slide and remo...

Page 19: ...p holder from the upper part refer to Figure 11 1 Remove the piston assembly O ring and the seals refer to Changing the Tip Ejector for Single Channel 2 Leave the piston exposed clean it with isopropa...

Page 20: ...MAN FP070105 Window 3digit Green MyPIPETMAN FP070106 Lubricant Tube for PIPETMAN 1g 5440011070 NOTICE For pipette service and maintenance or for extra spare parts please contact your local Gilson Serv...

Page 21: ...ilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipet...

Page 22: ...ilson Inc All rights reserved Specifications subject to change without notification errors excepted www gilson com contactus LT801622 D 2022 Gilson Inc All rights reserved Specifications subject to ch...

Reviews: