PI
P
E
T
M
A
N
®
D
IA
M
O
N
D
T
IP
S
14
PIPETMAN® DIAMOND TIPS
|
PIPETMAN® M
●
Mold and cavity references are marked on the collar, ensuring the traceability. For
quality assurance purposes, batch numbers appear on all packages (bags and boxes).
●
They form an air-tight seal with the tip holder, preventing the leaks that cause
poor accuracy and a lack of precision.
●
PIPETMAN DIAMOND Tips (except filter tips) may be autoclaved at 121°C for
20 minutes at 0.1 MPa.
NOTICE
Do not autoclave PIPETMAN Filter Tips as the filter will be damaged.
NOTE
To ensure the best performance from your Gilson pipette, you
shoud ALWAYS use PIPETMAN DIAMOND Tips (in accordance with
ISO8655) system, because PIPETMAN Tips were used to establish the
specifications.
PIPETMAN M MULTICHANNEL
P8x10M
D10
DL10
DF10
DFL10
0.5–10 µL
P12x10M
P8x20M
DL10
D200
DFL10
DF30
1–20 µL
P12x20M
P8x100M
D200
DF100
10–100 µL
P12x100M
P8x200M
D200
D300
DF100
DF200
DF300
20–200 µL
P12x200M
P8x300M
D200
D300
DF200
DF300
10–300 µL
P12x300M
P8x1200M
D1200
DF1200
50–1200 µL
P12x1200M
PIPETMAN M SINGLE CHANNEL
VARIABLE VOLUME MODELS
Model
PIPETMAN
DIAMOND Tips
Volume Range
P10M
D10*
DL10*
DF10
DFL10
0.5–10 µL
P20M
D200
DF30
2–20 µL
P100M
D200
D100
5–100 µL
P200M
D200
D300
DF200
DF300
20–200 µL
P300M
D200
D300
DF200
DF300
20–300 µL
P1200M
D1000
D1200
DF1000
DF1200
100–1200 µL
P5000M
D5000
500–5000 µL
P10mLM
D10mL
1–10 mL
Figure 7
Plastic adapter
Figure 8
P8x10M and P12x10M
Long tips
Short tips
Ejector spacer
Ejector spacer D10
* A plastic adapter is required to eject D10 and
DF10 tips (short tips). The adapter is supplied
with P10M. No adapter is required when using