1. Installations
Before installation of the MFPR-3100, please read the
“
Safety Precautions
”
of this manual.
1-1. Confirmation of Environment
This product operates properly within the temperature range of +0-35
℃
and the humidity range
of 10-90% (Non-condensing). If the temperature of this product is out of this range due to a
transportation or preservation, do not power on until the temperature reaches the specified range.
To prevent an over-heat, do not disturb the airflow of this product. Keep a certain space open, at
least 5 cm on the front and the rear of this product.
1-2. Assurance of Mains Voltage
Assure that the mains AC voltage to this product is within the range of 90-264V 50/60 Hz.
1-3. Connection of Power Cable and Power ON
The power cable must be connected properly to the AC IN connector on the rear panel.
In default, the system automatically power-on by connecting the AC power.
Also, the system power is fed by pushing the front power button.
After power on, cooling fans start rotating, operating system is booted within about 30 seconds,
then the system can be used.
1-4. Power OFF
Push the POWER switch on the front panel for shutting down the power. Also, the power can be
off from the operating system. Push the START button and click the
“shut down” button.