Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-123
3.19.6
Measured Offset Entry
A measurement can be saved and used as an offset. The command format for this purpose is:
Syntax:
[AP, BP, AR, BR, AD, or BD] OS DO EN
[AP, BP, AR, BR, AD, or BD] is a function code which describes the measurement that is to be
stored as an offset value. There are six possible function codes; they are interpreted as follows:
AP for the Sensor A power measurement
BP for the Sensor B power measurement
AR for the ratio measurement A/B
BR for the ratio measurement B/A
AD for the difference measurement A-B
BD for the difference measurement B-A
OS followed by DO indicates that the difference between the current offset and the current value
of the measurement described in the prefix, is to be saved as an offset value
EN is a required terminating suffix
Example:
AP OS DO EN
! SAVES MEASUREMENT A AS AN OFFSET
Example 2:
BP OS DO EN
! SAVES MEASUREMENT B AS AN OFFSET
Example 3:
AR OS DO EN
! SAVES MEASUREMENT A/B AS AN OFFSET
Example 4:
BR OS DO EN
! SAVES MEASUREMENT B/A AS AN OFFSET
Example 5:
AD OS DO EN
! SAVES MEASUREMENT A-B AS AN OFFSET
Example 6:
BD OS DO EN
! SAVES MEASUREMENT B-A AS AN OFFSET