Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-59
3.2.29
Halting Operation
ABORt
ABORt
Syntax:
ABORt
Example:
ABOR
! HALTS MEASUREMENT & TRIGGERING
Description:
This command stops operation, but it does not interrupt the completion of the current action.
For example, sensor calibration is not interrupted. Burst mode data collection is not interrupted.
3.2.30
Language Configuration
SYSTem:LANGuage NATIVE
SYSTem:LANGuage NATIVE
Syntax:
SYSTem:LANGuage NATIVE
Example:
SYST:LANG NATIVE
! SWITCHES ACTIVE GPIB COMMAND SET TO NATIVE LANGUAGE
Description:
Command modifies the active GPIB command set to native (8600) from SCPI. Query from is
not available for this command.