Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-31
3.2.16
Channel Configuration
CALCulate:POWer
CALCulate:RATio
CALCulate:DIFFerence
CALCulate?
CALCulate:UNIT
CALCulate:STATe
CALCulate:CrestFACtor:STATe
CALCulate:CrestFACtor:DATa?
CALCulate:PeakHOLd:STATe
CALCulate:PeakHOLd:DATa?
CALCulate:StatisicalFUNctions:STATe
CALCulate:StatisicalFUNctions:DATa?
CALCulate:POWer
Syntax:
CALCulate<channel 1 to 4>[:CHANnel]:POWer
space<
sensor 1 or 2>
Example:
CALC2:POW 1
! CONFIGURES CHANNEL 2 TO MEASURES SENSOR 1 POWER
Description:
This command configures a sensor to an individual channel, and the channel measures the
sensor power level. This command does not have a query format. For the example above, the
response to a CALC2? query is POW 1. This command should be executed only after the
designated sensor has been calibrated.
CALCulate:RATio
Syntax:
CALCulate<channel 1 to 4>[:CHANnel]:RATio
space
<sensor 1 or 2
comma
sensor 2 or 1>
Example:
CALC4:RAT 2,1
! CONFIGURES CHANNEL 4 TO MEASURE THE RATIO OF
! SENSOR 2 OVER SENSOR 1 POWER
Description:
This command configures a channel as a ratio of the power levels present at the two sensors.
When UNITs are DBM, RATio configuration produces measurements in dB. When UNITs are
Watts, RATio configuration measurements are percentage. This command does not have a
query format. For the above example, the response to a CALC1? query is RAT 2,1. This
command should be executed only after the designated sensor has been calibrated.
CALCulate:DIFFerence
Syntax:
CALCulate<channel 1 to 4>[:CHANnel]:DIFFerence
space
<sensor 1 or 2
comma
sensor 2 or 1>
Example:
CALC3:DIFF 2,1
! CONFIGURES CHANNEL 3 TO MEASURE
! SENSOR 2 MINUS SENSOR 1 POWER
Description:
This command configures a channel as the difference between the power levels present at the
two sensors. When UNITs are DBM, DIFFerence configuration produces measurements in
dBm. When UNITs are Watts, DIFFerence configuration measurements are watts. This
command does not have a query format. For the above example, the response to a CALC3?
query is DIF 2,1. This command should be executed only after the designated sensor has been
calibrated.