Series 8650A Universal Power Meters
1-6
Manual 31470, Rev. E, April 2001
1.2.4
The Rear Panel
The rear panels for the Models 8651A and 8652A are identical and are illustrated in Figure
-2. Any
options that have been installed in the unit will be noted on the serial number tag. Refer to the Special
Configurations section in the preface of this manual for detailed information about installed options or
other special configurations. Appendix C contains information on all available options for the 8650A.
Line Voltage Selection and Fuse
Select the correct operating voltage and install the proper fuse in this housing. Refer to Section 1.2.2 for
instructions and precautions on how to select the voltage and replace the fuse.
Inputs/Outputs
Four BNC-type connectors interface the 8650A to other equipment
•
Trigger Input
accepts a TTL input for triggering and gating measurements as defined by the
Gate/Trigger menu selections, or under GPIB control. Maximum input without damage is 15 V.
•
V
PROP
F In
accepts a voltage input that is proportional to frequency and causes the 8650A to
apply appropriate frequency-related cal factors. Maximum input without damage is 15 V.
•
Analog Out A and Analog Out B
each provide an output voltage that is proportional to
the measured power level of the respective sensors connected to the front panel.
Remote Interface
•
GPIB
is a DB-24 connector to interface the 8650A to a host computer over the GPIB.
•
RS-232
is a DB-9 connector for interfacing the meter with serial communication equipment.
Figure 1-2: The 8650A Rear Panel
120Vac
Fuse
110-120V
T250 .50A
220-240V
T250 .35A
-- Line
25VA
MAX
U.S. Patent 4,794,325
OPTION 01
WARNING
WARNING
For continued fire protection
replace fuse with same
type and rating
No operator serviceable parts
inside. Refer servicing to
service trained personnel
LINE VOLTAGE
SELECTION
GPIB
RS-232
Trigger
Input
Analog
Out A
Analog
Out B
V
In
F
∞
.
!
A
B
Cal
NOTE:
The Sense In connectors will be relocated to the rear panel when Option 03 (Model
8651A) or Option 04 (Model 8652A) is installed. The Sense In and the Calibrator Out
connectors will be relocated to the rear panel when Option 13 (Model 8651A) or Option 14
(Model 8652A) is installed (descriptions of these options are in Appendix C).
☛
☛
☛
☛
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...