Series 8650A Universal Power Meters
3-82
Manual 31470, Rev. E, April 2001
3.12.2
Frequency
Cal factors are stored in the sensor’s EEPROM by frequency. Specifying a frequency causes the 8650A to
apply the cal factor appropriate to that frequency. To cancel the use of cal factors, specify a frequency of
50 MHz (this is the frequency of the front panel Calibrator reference output, and has a cal factor of
zero).
Commands which specify a frequency are based on the FR function. The command format is:
Syntax
:
[AE or BE] FR [n] [HZ or KZ or MZ or GZ]
[AE or BE] prefix specifies Sensor A or Sensor B.
FR specifies a frequency value.
[n] specifies the frequency value (the units are Hz, kHz, MHz, or GHz, depending on the
terminating suffix used).
A terminating suffix is required (HZ, KZ, MZ, or GZ).
Examples
:
OUTPUT 713;AE FR 5.67 GZ
! frequency for sensor A is 5.7 GHz
OUTPUT 713;AE FR 1.0E9 HZ
! frequency for sensor A is 1E9 Hz (1 GHz)
OUTPUT 713;BE FR 84.6 MZ
! frequency for sensor B is 84.6 MHz
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...