Calibration and Testing
Manual 21568, Rev. F, March 2008
50 MHz, -30 dBm
Pulse output settings:
PRF 100 Hz
Width: 2 msec
Pulse State: ON
NOTE:
Applies to 80350A, 80353A and 80354A sensors only.
1.
Set the Pulse Sensor trigger to Internal and sample delay to 1 ms.
2.
Set the sensor trigger level to -5 dBm. Set the pulse amplitude of the RF source to 0 dBm
and verify triggering by noting that the sample delay pulse is present
3.
Set the RF source level to -20 dBm verify triggering does not occur.
4.
Set the sensor trigger level to -10 dBm and set the RF source level to 0 dBm and verify
trigger operation.
5.
Set the Pulse Sensor to external trigger (EXT TRIG). Set the sensor external trigger level to
1.7 V
DC
. Verify triggering with +5 V pulse input. Set the input level to 0 dBm. Verify that the
sensor is triggered and the meter displays a valid power reading.
6.
Disable the external trigger then set the input level to -10 dBm. Verify that the sensor is not
triggered. The peak sensor should not trigger.
4.4.2.2
High Power Peak Sensors
Set the RF source to the following settings:
50 MHz, -30 dBm
Pulse output settings:
PRF 100 Hz
Width: 2 msec
Pulse State: ON
NOTE:
Applies to 80351A, 80352A or 80355A sensors only.
1.
Remove the attentuator before connecting the sensor to the pulse source.
Do not
23 dBm peak input power to the sensor.
2.
Set the Pulse Sensor trigger to Internal and sample delay to 1 ms.
3.
Select the appropriate table on the datasheet based on the model number of the sensor
being tested.
4.
Set the sensor trigger level to +15 dBm for the 80351A sensor and +25 dBm for the
80352A and 80355A sensors. Set the syntehsizer level to 0 dBm and verify trigger
operation.
5.
Set the RF source level to -20 dBm and verify triggering does not occur.
6.
Set the trigger level to +10 dBm for the 80351A sensor and +20 dBm for the 80352A and
80355A sensors. Set the RF source level to 0 dBm and verify trigger operation..
7.
Set the Pulse Sensor to external trigger (EXT TRIG). Set the external trigger to 1.7 V
DC
.
Verify triggering with +5 V pulse input. Set the input level to 0 dBm. Verify that the sensor
is triggered and the meter displays a valid power reading.
8.
Disable the external input trigger then set the RF source level to -10 dBm. Verify that the
sensor is not triggered. The peak sensor should not trigger.
NOTE:
Reset the sensor trigger when testing for a non-triggered state after the sensor has
been triggered. Resetting the sensor trigger can be acheived by varying the sample delay by
Summary of Contents for 80350A
Page 4: ......
Page 10: ...Series 8035XA Peak Power Sensors vi Manual 21568 Rev F March 2008 ...
Page 13: ...Manual 21568 Rev F March 2008 ...
Page 16: ...Series 8035XA Peak Power Sensors xii Manual 21568 Rev F March 2008 ...
Page 17: ...Manual 21568 Rev F March 2008 ...
Page 18: ...Series 8035XA Peak Power Sensors xiv Manual 21568 Rev F March 2008 ...
Page 24: ...Series 8035XA Peak Power Sensors 1 6 Manual 21568 Rev F March 2008 ...
Page 90: ...Series 8035XA Peak Power Sensors 6 22 Manual 21568 Rev F March 2008 ...
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......