4 | The product at a glance
G 1501
10 / 37
B-H86.0.12.DB2-2.1
Up / Down button
Press briefly
Display of the min/max value
Change value of the selected parameter
Long press
Reset the min/max value of the current measure-
ment
Both simultaneously
Rotate display, overhead display
Function key
Press briefly
Freeze measurement (Hold)
Return to measurement display
Call up next parameter
Long press, 2s
Start menu configuration,
(ONF
appears in the dis-
play
Long press, 4s
Start automatic calibration,
(AL
appears in the dis-
play
4.4
Connections
BNC connection
Connection for pH electrode
Un/locking with rotating ring on the cable plug
2x 4mm banana
Connection for temperature sensor or reference
electrode
CAUTION
Ensuring water tightness!
The product guarantees protection from spray water, rain or accidental immersion in
water. This protection for the plug connector is only guaranteed when plugged in.
Moisture or contaminants on the contacts can result in incorrect measurement results.
– Protect contacts from soiling and moisture!
– Dry off damp plug connectors as quickly as possible!
NOTE
The temperature measurement can be influenced by conductive liquids on the banana
sockets. We recommend always keeping the connections dry.