Measuring:
When exceeding the maximum range limit (10kHz) independently from the limit setting an error
message will be displayed (“Err.1“).
6.5. Measuring of rotation speed (TTL, switching-contact)
This chapter describes how to configure the device for measuring rotation speed.
This instruction demands that you se
lected “
rPn
“ as input type like it is explained in chapter 6.1.
The display must show
INP
.
Parameter
Button 1
Value
Button 2 / 3
Notes
SENS
TTL
TTL-signal
NPN
Switching contact, NPN
For direct connection of a passive switching contact (e.g. push
button, relay) or transmitter with NPN output.
There is a internal pull-up-resistor
PNP
Switching contact, PNP
For direct connection of a transmitter with PNP output.
There is a internal pull-down-resistor.
D,V
1 ... 1000
divisor
DP
"- - - -"
"- - - . -"
"- - . - -"
"- . - - -"
Decimal point
(see info decimal point)
O
V
TP
Output configuration will be explained in chapter 7. According to
your settings different parameters will be shown.
Finish configuration:
Press
Button 1
as often till the device executes a reset. The values will be stored at it.
.
Decimal point:
Use the decimal point position to change the resolution of your measurement. The more the deci-
mal point position is on the left, the finer the resolution will become. Please note that you lower the
maximum value that can be displayed, either
Example:
50 rotations per minute.
without decimal point: display = 50, the maximum value 9999 rotations per minute.
decimal point on the left e.g. XX.XX: display = 50.00, maximum value 99.99 rotations per minute.
Connection:
For the connection of a frequency-transmitter, please follow the instructions given in chapter 5.3.6.
When connecting a switching-contact-transmitter with increased frequency range (= with external
circuitry) you have to select TTL as input signal.