8  Alarm functions 

There are 3 possible settings for the alarm function: off (AL oFF), on with horn (AL on) and on 
without horn (AL no.So) 
The alarm will be issued under the following circumstances: 

measuring value higher or equalling upper alarm limit ’AL.Hi’ 

- sensor error 
- measuring values exceeding/falling below measuring ranges 

battery voltage too low or error in device (’Err.7’) 

The alarm is supported by the interface, thus, it can be monitored by a connected computer. 
The acoustic alarm is depending of the value. It is an interrupting tone, increasing in intensity with 
the value, above 300ppm the tone is permanent. 
Visual alarming: „AL.Hi“ flashes in the display 
Additionally the exceeding of the MAK value is signalled with the MAK-arrow independently from 
the alarm settings. 
Because the maximum working place concentration differ from country to country, this value is 
adjustable. 

9  Adjustment 

The adjustment of the GCO100 is stable within 5% within in a year. 
The measuring precision can be checked by means of suitable test-gas mixtures and the suitable 
testing accessories. Alternatively the instrument can be sent to the manufacturer for test and 
recalibration. 

Attention! If the adjustment is altered, the included Calibration protocol is no more valid! 

 

9.1  Adjustment menu 

To adjust a measuring offset and scale proceed like follows: 

  Switch off instrument. 

  Switch on  

and press 

“Hold”-key during the segment test

 (                    ), until the display shows the first 

parameter  “OFFS”  

 

Press “up” or “down” key, the currently selected offset adjustment appears. 

 

Choose the desired value by pressing “up” or “down” key. (max. selectable values: ±20ppm) 

  Enter by pressing On/Off-key: OFFS appears in the display again 

  Selsect the next parameter by pressing On/Off-key: SCAL appears in the display 

 

Press “up” or “down”  key, the currently selected scale appears  

 

Choose the desired value by pressing “up” or “down” key. (28.00 ... 62.00 nA/ppm). 

  Store the values by 

pressing ‘On/Off’ key. The instrument will restart (segment test). 

Please note: 

If during the changing of the offset adjust no key is pressed within 120 
seconds, the input will be aborted. Eventually made changes won’t be 
stored! 

 

9.2  Automatic zero point adjusting 

To compensate a deviation of the zero point proceed like following: 

  Switch on instrument and bring it to a CO-free atmosphere (fresh air or good vented room 

without any CO sources) 

  If there was measured at higher CO levels before, please wait at least 3 minutes. 

  press 

Hold-key 5 seconds

, until the display shows “

nuLL

”   

The zero point adjusting is now performed automatically and will be stored in the instrument. The 
instrument will automatically show the measured value after successful adjustment.  
If the adjusting could not be performed (display shows nuLL permanently), there is a higher CO-
value in the ambient air or the sensor is defect. Turn device off and on to return to measuring 

– 

Note: zero point adjusting was not stored.

 

The zero point can be viewed in the adjustment menu 

at parameter “OFFS”. 

 

 

Summary of Contents for GREISINGER GCO 100

Page 1: ... 49 9402 9383 0 info greisinger de www greisinger de Operating Manual Carbon Monoxide CO Measuring Instrument with alarm function as of version 1 4 GCO 100 Please carefully read these instructions before use WEEE Reg Nr DE 93889386 Please consider the safety instructions Please keep for future reference ...

Page 2: ...HE SERIAL INTERFACE 11 11 1 INTERFACE 11 11 2 SUPPORTED INTERFACE FUNCTIONS 11 12 SPECIFICATION 12 1 General Advice Read through this document attentively and make yourself familiar to the operation of the device before you use it Keep this document in a ready to hand way in order to be able to look up in the case of doubt The manufacturer will assume no liability or warranty in case of usage for ...

Page 3: ...gnificant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This symbol point out processes which can indirectly influence operation or provoke unforeseen reactions at non observance 2 3 Safety guidelines This device has been designed and tested in accordance with...

Page 4: ...swallowing give large volumes of water DO NOT induce vomiting Obtain medical treatment 3 Product Specification 3 1 Scope of supply Device GCO 100 incl 9V battery block Operation manual Calibration protocol 3 2 Operating and maintenance advice Battery operation If bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a ...

Page 5: ...HLD Measured value is frozen key 3 4 2 Pushbuttons Left key on off button for power off press longer Middle key Max display of the maximum measured value MAX arrow in display Right key Hold hold frozen of the current measure value HLD in display 4 3 Connections 1 Interface Connect to optically isolated interface adapter accessory GRS 3100 USB3100 2 Sensor opening The mains socket is located at the...

Page 6: ... the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 5 2 Disposal instructions Batteries must...

Page 7: ...pm Strong smoker 20ppm Like with alcohol concentration in blood also the CO is decomposed in the blood around every 5h the CO content is cut by half Via the CO concentration of exhaled air the saturation of the CO in blood haemoglobin can be derived The carboxy haemoglobin in can be displayed directly by the instrument COHb p r t Configuration of the instrument This value is an estimation of the c...

Page 8: ...nit mg m3 Selection of the display unit COHb AL Alarm function OFF No alarm ON Alerting over display and buzzer factory setting NO SO Alerting only via display AL K Alarm boundary not available if alarm function is deactivated AL oFF 0 1000 Alarm boundary in the selected display unit factory setting 10ppm RAIL Warning boundary for maximum working place concentration 0 100 Value for warning boundar...

Page 9: ...ment Switch on and press Hold key during the segment test until the display shows the first parameter OFFS Press up or down key the currently selected offset adjustment appears Choose the desired value by pressing up or down key max selectable values 20ppm Enter by pressing On Off key OFFS appears in the display again Selsect the next parameter by pressing On Off key SCAL appears in the display Pr...

Page 10: ... test gas cap GZ 10 to the instrument Connect the test gas cap GZ 10 to the valve unit via the hose The Gas bottle has to stand upright for the flow display of the gas valve unit is able to work properly Scale Adjustment Keep bottle upright open valve slowly until the ball in the flow indicator shows 0 5 l min lowest marking of indicator Wait until the measurement display at has stabilised ca 1min...

Page 11: ...ng of the device e g Material selection EBS 20M 60M 20 60 channel software to display the measuring values In case you want to develop your own software we offer a GMH3000 development package including a universally applicable Windows functions library GMH3000 DLL with documentation that can be used by the most programming languages Suitable for Windows XP Windows Vista Windows 7 Windows 8 8 1 Win...

Page 12: ...time 0 90 RH Storage temperature 10 50 C Housing impact resistant ABS membrane keyboard transparent panel Front side IP65 integrated pop up clip for table top or suspended use Dimensions 142 x 71 x 26 mm L x W x D Wight approx 155 g Power supply 9V battery in scope of supply as well as additional d c connector diameter of internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable pow...

Reviews: