INGLE ACTUATOR DOES NOT WORK
.
1.
Motor-plug is not connected
properly.
1.
Connect plug properly.
2.
MCP is defective.
2.
MCP must be replaced (please
contact the customer service).
3.
Control-box is defective.
3.
Control-box must be replaced
(please contact the customer
service).
4.
Motor-cable or plug is defective.
4.
Cord / plug must be replaced (please
contact the customer service).
5.
Motor is defective.
5.
Motor must be replaced (please
contact the customer service).
If the treatment table shows a malfunction not noted on the schedule above or if
you have checked all items on the above schedule but the problem still exists,
please contact the customer service (s. chapter 14.1 R
EPAIR
).
14.1 Repair
Y
OU CAN CONTACT THE
G
HARIENI CUSTOMER SERVICE USING THE
FOLLOWING CONTACT INFORMATION
:
C
USTOMER SERVICE HOTLINE
:
+49 (0)2841/88300-40
E-
:
❖
We are responsible for safety and reliability of the SPL med under the following
conditions only.
1.
When contacting the customer service, the information about the
serial number
is absolutely necessary.
2.
Repairs and modifications must be carried out by Gharieni Group GmbH or a
person authorized by Gharieni Group GmbH.
3.
The electrical installation of the room the SPL med is installed in must meet the
current legal regulations at the site where the SPL med is operated.
4.
The SPL med is operated as described in the instruction manual.
5.
Unauthorized modifications or manipulating the SPL med will void the liability
of the manufacturer.
6.
All electric components are designed for continuous operation. Leave the
SPL med connected when it is used every day.
7.
Check the power cord regularly for signs of damage.
8.
When the power cable is damaged, it must be replaced by the manufacturer or an
authorized aftersales service or a qualified professional only.