background image

*3-9900.094*

For complete product information, please download the full 
9900 Transmitter manual at www.gfsignet.com

Signet 9900 H COMM Module Instruction Sheet

3-9900.094   Rev. 4   05/18

The H COMM Module enables communication between the 9900 and 
a HART

®

-enabled device.  The HART (

H

ighway 

A

ddressable 

R

emote 

T

ransducer) Protocol superimposes digital signals on top of the 4 to 20 mA 

analog signal. 

NOTE:

 With H COMM Module installed, a minimum of 24 V is required for 

loop-powered systems.

English

Installation

If the 9900 Base Unit will be mounted in a panel, plug-in modules may be installed either before or after the base unit is mounted.

If the 9900 Base Unit will be mounted using the accessory wall mount kit (3-9900.392), install plug-in modules 

fi

 rst. 

If the Direct Conductivity/Resistivity Module or the 4 to 20 mA Output Module will be included in your unit, install the HCOMM 
module 

fi

 rst, then install the other module over the H COMM module

To install the H COMM module, carefully align the module pins into its plug (see illustration) and push the module straight in until the 
tabs on the bottom edge snap into place.

To uninstall, squeeze tabs, grasp the module and pull straight out.

Wiring

3-9900-1P Panel Mount

3-9900.395

H COMM Module

DC Power

Loop V

oltage

+

+

PLC

Device

Optional ammeter

(recommended for

4 to 20 mA trim only)

Power

Supply

9900 Device
Connections

Loop +

Loop –

(Primary HART Master)

Hand-held device

(Secondary HART Master)

Red

Black

Connecting HART with a Loop-powered sensor

CAUTION

Exercise care when installing module.

 Do not bend connecting pins. 
Align pins and connectors then push module firmly 
into place.

Avoid Electrostatic Discharge (ESD)

 

 Minimize handling of module to reduce the 
possibility of damage due to ESD.

 

Handle module by the edges.

 

Never touch any exposed circuitry or contacts.

 

 Wear an anti-static wristband, stand on an anti-
static mat, or keep one hand touching a properly 
grounded pipe or other properly grounded piece 
of metal when handling module.

• 

English

• 