1

Installation

Flusher disinfector S-406

9904

Flusher

disinfector

S-406

Serial No. 203.238 -

4996959-01

INSTALLATION

0301

Produced: 1998

Summary of Contents for S-406

Page 1: ...1 Installation Flusher disinfector S 406 9904 Flusher disinfector S 406 Serial No 203 238 4996959 01 INSTALLATION 0301 Produced 1998...

Page 2: ...4 Fixation in wall __________________________________________ 4 Connection of electricity water drain and possibly steam ________ 6 Goods arrangement chart __________________________________ 7 Deterge...

Page 3: ...ncerned must study valid handbooks and service manuals prior to any repairs or service work The machine may not be sprayed with water Caution must be observed when using corrosive detergents Safety pr...

Page 4: ...infector may be placed in a free standing position on four legs or mounted on a wall in a special wall bracket To facilitate cleaning a lateral distance of 100 mm to an adjacent wall or other equipmen...

Page 5: ...d remove the transport stay and attach the legs before fixing the disinfector in position Make sure that the disinfector hooks properly into the bracket when it is mounted on the wall fitting V300 Fig...

Page 6: ...l box can be unscrewed and pulled forward to facilitate installation Fig 4 Electrical connection see electric circuit diagram in Service instructions To prevent clogging of filters and valves flush th...

Page 7: ...art Place the notice with user s instruction on the wall above the disinfector Make sure that the lid does not obscure the chart Fig 5 Connection of water and steam supply Connect the disinfector to d...

Page 8: ...in the cans can be corrosive The container may also be placed on the floor or on the lower shelf of an adjacent sink unit up to a distance of about 70 cm from the disinfector The bottom of the contain...

Page 9: ...ld enter the service program as follows Simultaneously press 4 and 5 for 10 seconds The machine is then reset to the Service mode see also under the heading Service program The push buttons 1 5 will h...

Page 10: ...nt dosage systems two for detergent dosage and one for descaling compound For descaling and spray dosage the amount of compound can be set in the service program see Illustration 8 Set by the service...

Page 11: ...ector the lid shall open again The limit for this safety precaution is between 6 and 12 mm as shown in figure 9 Use the accompanying spacer bar to check this function Stand the bar on edge so that it...

Page 12: ...and apparatus with steam generator The final temperature is higher than 90 C The PROCESS COMPLETE lamp lights when the program is over Fig 10 Indicating lamps L M K Process in operation Process compl...

Page 13: ...0 C Noise level ca 59 dB A Power Model Electrical connection Fuse requirement L1 230 V 1N PE 50 Hz TN 1x16 A 2 94 kW L2 230 V 3 PE 50 Hz 3x16 A 3 91 kW L3 400 V 3N PE 50 Hz TN 3x10 A 3 91 kW L5 240 V...

Page 14: ...f no steam pressure is specified the disinfector nozzle is delivered with an orifice diameter of 1 8 mm If the pressure used is less than 300 kPa the nozzle orifice should be made larger as shown in t...