Maintenance
4–4
Weekly
CLEAN THE SPRAY JETS AND HEADERS
1.
With power turned off, open Chamber Door(s)
(a flashlight may be
needed)
.
2.
Push the debris through the jet and into the header with a small wire.
3.
Remove the flush plugs from the end of the spray headers.
4.
In manual mode, fill sump with water and turn pump on for 30 seconds.
5.
Reinstall flush plugs
6.
Turn on gravity drain and allow to drain totally.
INSPECT THE TEMPERATURE AND WATER LEVEL PROBES FOR DEBRIS
AND DAMAGE
1.
Locate the temperature and water level probes in the wash and rinse
section sumps.
(Accessed externally)
2.
Drain wash and rinse tanks.
3.
Remove sump cover for each tank.
4.
Gently wipe off any debris around temperature and water level probes
with a cloth.
5.
Replace sump cover for each tank.
MAINTAIN THE CABINET
To maintain cabinet luster, wipe down with non-abrasive stainless steel
cleaner in straight line with direction of grain.